Suppr超能文献

使用激光多普勒血流仪测量人体牙髓血流:一种在手术操作过程中能够实现牙髓血流测量稳定性和可重复性的技术。

Measurement of pulpal blood flow in humans using laser Doppler flowmetry: a technique allowing stability and repeatability of pulpal blood flow measurement during surgical manipulations.

作者信息

Anderson K K, Vanarsdall R L, Kim S

机构信息

Department of Orthodontics, University of Pennsylvania School of Dental Medicine, Philadelphia 19104, USA.

出版信息

Int J Adult Orthodon Orthognath Surg. 1995;10(4):247-54.

PMID:9082013
Abstract

This study was undertaken to validate the repeatability of pulpal blood flow in human teeth with the laser Doppler flowmetry technique using a repeatable method of attachment of the laser Doppler probe. The variation in laser Doppler flowmetry measurements was evaluated over time periods that ranged from hours to days. In a prospective study, the pulpal blood flow was recorded from a maxillary central incisor (randomly selected) at a site directly gingival to the orthodontic bracket of three subjects. Measurements were made eight times over a 2-week period and every 30 minutes over 4 hours for eight recordings. The results indicated that there was only a slight difference between the temporal variability of the laser Doppler measurements over the course of 4 hours (intraday variability) or 2 weeks (interday variability). Furthermore, it appeared that subjects varied in their mean value recordings and may have varied widely in their respective signal-to-noise ratios.

摘要

本研究旨在通过使用激光多普勒探头的可重复附着方法,采用激光多普勒血流仪技术验证人牙齿牙髓血流的可重复性。在数小时至数天的时间段内评估激光多普勒血流仪测量值的变化。在一项前瞻性研究中,在三名受试者正畸托槽正牙龈侧的部位,从一颗上颌中切牙(随机选择)记录牙髓血流。在2周内进行8次测量,在4小时内每30分钟进行一次测量,共进行8次记录。结果表明,在4小时(日内变异性)或2周(日间变异性)的过程中,激光多普勒测量的时间变异性之间只有轻微差异。此外,受试者的平均值记录似乎有所不同,并且各自的信噪比可能差异很大。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验