Antretter H, Bonatti J, Cottogni M, Hangler H, Müller L C, Mair P, Dapunt O E
Abteilung für Herzchirurgie, I. Chirurgischen Universitätsklinik Innsbruck.
Acta Med Austriaca. 1996;23(5):159-64.
Energy delivery (stimulation amplitude) of a pacemaker should be adjusted to the depolarisation threshold of the heart, which is a nonconstant magnitude. As programming of the stimulation amplitude is time-consuming, many pacemakers are never adjusted to the measured threshold referring to energy delivery. The unnecessary energy loss subsequently reduces the longevity of the implanted devices and remains the weak point of modern pacemakers. Autocapture function (automatic stimulation threshold search, subsequently automatic amplitude regulation 0.3 V above measured threshold, evoked response detection to verify the efficacy of every stimulus and back up pulse in case of loss of capture) is for the first time realized in an extremely small, rate adaptive, multiprogrammable single chamber pacemaker (Microny SR+ 2425T). The pulse generator was tested in a multicentric, european clinical investigation study. These pacemakers were implanted in 8 patients (5 male, 3 female) between December 1994 and April 1995. Automatic measurement of Autocapture threshold revealed 0.7 +/- 0.15 V at implantation, 1.5 +/- 0.3 V after 1 month, 1.3 +/- 0.14 V after 6 months and 1.3 +/- 0.14 V 1 year after implantation. A safety margin of 0.3 V is added to the threshold value to secure capture, therefore mean stimulation amplitude is 1.6 V after 12 months implantation time. The Autocapture feature minimizes current drain and that means that this extremely small generator provides a comparable life as the twice as large common generators used at present time.
起搏器的能量输送(刺激幅度)应根据心脏的去极化阈值进行调整,该阈值并非恒定不变。由于刺激幅度的编程耗时,许多起搏器从未根据测量到的能量输送阈值进行调整。随后不必要的能量损耗会缩短植入设备的使用寿命,这仍是现代起搏器的薄弱环节。自动夺获功能(自动刺激阈值搜索,随后自动将幅度调节至高于测量阈值0.3V,检测诱发反应以验证每次刺激的有效性,并在夺获失败时提供备用脉冲)首次在一款极小的、频率自适应的、多程控单腔起搏器(Microny SR+ 2425T)中得以实现。该脉冲发生器在一项欧洲多中心临床研究中进行了测试。1994年12月至1995年4月期间,这些起搏器被植入了8名患者体内(5名男性,3名女性)。自动夺获阈值的测量结果显示,植入时为0.7±0.15V,1个月后为1.5±0.3V,6个月后为1.3±0.14V,植入1年后为1.3±0.14V。在阈值基础上增加0.3V的安全裕度以确保夺获,因此植入12个月后的平均刺激幅度为1.6V。自动夺获功能可将电流消耗降至最低,这意味着这款极小的发生器与目前使用的体积两倍大的普通发生器具有相当的使用寿命。