Suppr超能文献

自动阈值跟踪在预计起搏器寿命及起搏相关成本方面带来的益处。

Benefits in projected pacemaker longevity and in pacing related costs conferred by automatic threshold tracking.

作者信息

Boriani G, Biffi M, Branzi A, Mininno A, Sigliano R

机构信息

Institute of Cardiology, University of Bologna, Policlinico S. Orsola, Via Massarenti, 9, 40138 Bologna, Italy.

出版信息

Pacing Clin Electrophysiol. 2000 Nov;23(11 Pt 2):1783-7. doi: 10.1111/j.1540-8159.2000.tb07018.x.

Abstract

The aim of this study was to calculate the battery current drain of different VVI-VVIR pulse generators, with and without automatic adjustment of stimulation output, from the same manufacturer, and obtain projections on pacemaker longevity and pacing related costs. Twenty-five patients who had received a VVIR pacemaker (Microny SR+ 2425T) with Autocapture function, an algorithm able to provide automatic adaptation of device output coupled with continuously monitored capture threshold, were analyzed. Calculations of current drain and projections of device longevity were based on diagnostic data retrieved from device memory and on comparisons of multiple programmed settings of the same device (Microny SR+ 2425T with Autocapture, Low Output [output = 2 x Autocapture threshold, but not less than 2.4 V] versus Standard Output [5 V]). In addition, devices with Autocapture but with different battery capacities (0.35 Ah and 0.79 Ah, Microny and Regency, respectively) were compared to a standard device (battery capacity = 0.78 Ah) at 2.5-V and 5-V output settings. According to a series of assumptions, Autocapture increased device longevity by 53% (Microny) and by 245% (Regency) over a 15-year follow-up with a consequent reduction in pacing related costs of 25% (Microny) and of 57% (Regency) compared to the standard settings of a VVIR device from the same manufacturer. In conclusion, pacing with constant adaptation of generator output coupled with continuous monitoring of stimulation threshold has the potential to increase generator longevity and to reduce pacing related costs as compared to a standard device from the same manufacturer.

摘要

本研究的目的是计算同一制造商生产的、具备和不具备刺激输出自动调整功能的不同VVI-VVIR脉冲发生器的电池电流消耗,并得出起搏器使用寿命和起搏相关成本的预测。对25名接受了具有自动夺获功能(一种能够根据持续监测的夺获阈值自动调整设备输出的算法)的VVIR起搏器(Microny SR+ 2425T)的患者进行了分析。电流消耗的计算和设备使用寿命的预测基于从设备内存中检索到的诊断数据,以及同一设备(具有自动夺获功能的Microny SR+ 2425T,低输出[输出 = 2×自动夺获阈值,但不低于2.4 V]与标准输出[5 V])的多个程控设置的比较。此外,将具有自动夺获功能但电池容量不同(分别为0.35 Ah的Microny和0.79 Ah的Regency)的设备在2.5 V和5 V输出设置下与标准设备(电池容量 = 0.78 Ah)进行了比较。根据一系列假设,在15年的随访中,与同一制造商的VVIR设备的标准设置相比,自动夺获功能使设备使用寿命延长了53%(Microny)和245%(Regency),同时起搏相关成本降低了25%(Microny)和57%(Regency)。总之,与同一制造商的标准设备相比,持续调整发生器输出并持续监测刺激阈值的起搏方式有可能延长发生器使用寿命并降低起搏相关成本。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验