Gawalko E J, Nowicki T W, Babb J, Tkachuk R, Wu S
Canadian Grain Commission, Grain Research Laboratory, Winnipeg, MB, Canada.
J AOAC Int. 1997 Mar-Apr;80(2):379-87.
A method is described for the determination of Pb, Cd, Cu, and Se in cereal samples. An atomic absorption spectrometer equipped with a transverse-heated graphite furnace with Zeeman background correction was used for all determinations. Sample preparation was performed by closed-vessel microwave digestion using nitric acid and focused openvessel microwave digestion using nitric acid-hydrogen peroxide. Both techniques were evaluated by using 15 cereal reference materials and comparing results with certified or reference values for each element. Cereal reference standards obtained from the Community Bureau of Reference (Europe), the National Institute of Standards and Technology (USA), the National Institute for Environmental Studies (Japan), the National Research Centre for Certified Reference Materials (People's Republic of China), and the Canadian Grain Commission were used. Application of a series of t-tests, conducted according to Sidak's modified Bonerroni t-procedure, showed that both techniques yielded accurate results for cereal reference materials. Some differences from certified and reference values, however, were found for each element.
描述了一种测定谷物样品中铅、镉、铜和硒的方法。所有测定均使用配备横向加热石墨炉和塞曼背景校正的原子吸收光谱仪。样品制备采用硝酸密闭容器微波消解和硝酸 - 过氧化氢聚焦开放容器微波消解。通过使用15种谷物标准物质并将结果与每种元素的认证值或参考值进行比较,对这两种技术进行了评估。使用了从欧洲共同体参考局、美国国家标准与技术研究院、日本国立环境研究所、中国国家认证认可监督管理委员会国家计量科学研究院和加拿大谷物委员会获得的谷物参考标准。根据西达克修正的邦费罗尼t检验程序进行的一系列t检验表明,这两种技术对谷物标准物质均产生了准确的结果。然而,每种元素都发现了与认证值和参考值的一些差异。