Abstract

Two procedures for the determination of serum and urinary creatinine by first generation AutoAnalyzer have been positively tested. These methods enable to perform these analyses by cam 60 and 100 respectively, in order to make the influence of carryover neglectable. Very regular peaks with clear-cut resolution can be obtained. The range of concentrations which can be reliably evaluated corresponds to 0,5-10,0 mg/100 ml for serum and to 20-500 mg/100 ml for urine, for this analysis neither diluition nor the dialysis of the sample are required.

摘要

已对第一代自动分析仪测定血清和尿肌酐的两种方法进行了阳性测试。这些方法分别能够通过cam 60和100进行这些分析,以使残留影响可忽略不计。可以获得分辨率清晰的非常规则的峰。对于该分析,血清中可可靠评估的浓度范围为0.5 - 10.0 mg/100 ml,尿液中为20 - 500 mg/100 ml,此分析既不需要样品稀释也不需要透析。