快速、灵敏的高效液相色谱法测定环孢素A及其代谢物M1、M17和M21。

Rapid, sensitive high-performance liquid chromatographic method for the determination of cyclosporin A and its metabolites M1, M17 and M21.

作者信息

Khoschsorur G, Semmelrock H J, Rödl S, Auer T, Petek W, Iberer F, Tscheliessnigg K H

机构信息

Department of Laboratory Medicine, University of Graz, Austria.

出版信息

J Chromatogr B Biomed Sci Appl. 1997 Mar 7;690(1-2):367-72. doi: 10.1016/s0378-4347(96)00364-7.

Abstract

Cyclosporin A (CyA) and its metabolites seem to have nephro-, hepato- and neurotoxic side effects. Immunosuppressive therapy is a narrow path between the risk of rejection by underimmunosuppression and toxic organ damage by overdosage. Thus CyA dosage must be calculated to avoid the risks of organ rejection through underdosage and toxic organ damage through overdosage or accumulation of metabolites. In routine monitoring of CyA therapy, it can be important to measure not only the parent drug but also the metabolites. We describe a rapid and isocratic high-performance liquid chromatographic method for measurement of CyA and its metabolites M1, M17 and M21 in whole blood. CyA was detected by ultraviolet absorption at 212 nm with a CN analytical column maintained at 50 degrees C and recycling of hexane-isopropanol as mobile phase for improved long-term column stability and efficiency. The minimum detectable concentration of CyA and the three metabolites was 10 ng/ml blood. Our modified HPLC method for the determination of CyA and its metabolites is a simple (isocratic), rapid (the retention times were 7.1 min for CYD, internal standard, 8.9 min for CyA, 11.0 min for M21, 12.9 min for M17 and 16.3 min for M1) and economical method suitable for measuring the concentration of the major metabolite, M17, and for routine monitoring of CyA-treated patients.

摘要

环孢素A(CyA)及其代谢产物似乎具有肾毒性、肝毒性和神经毒性副作用。免疫抑制治疗是一条狭窄的路径,介于免疫抑制不足导致的排斥风险和过量用药导致的器官毒性损伤之间。因此,必须计算CyA的剂量,以避免因剂量不足导致器官排斥以及因过量用药或代谢产物积累导致器官毒性损伤的风险。在CyA治疗的常规监测中,不仅测量母体药物,还测量代谢产物可能很重要。我们描述了一种快速等度高效液相色谱法,用于测量全血中的CyA及其代谢产物M1、M17和M21。CyA通过在212nm处的紫外吸收进行检测,使用在50℃下保持的CN分析柱,并以己烷 - 异丙醇作为流动相进行循环,以提高长期柱稳定性和效率。CyA和三种代谢产物的最低检测浓度为10ng/ml血液。我们改进的用于测定CyA及其代谢产物的高效液相色谱法是一种简单(等度)、快速(内标CYD的保留时间为7.1分钟,CyA为8.9分钟,M21为11.0分钟,M17为12.9分钟,M1为16.3分钟)且经济的方法,适用于测量主要代谢产物M17的浓度以及对接受CyA治疗的患者进行常规监测。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验