Reproducibility and reliability of volumetric measurements of olfactory eloquent structures.

作者信息

Yousem D M, Geckle R J, Doty R L, Bilker W B

机构信息

Department of Radiology, University of Pennsylvania Medical Center, Philadelphia 19104, USA.

出版信息

Acad Radiol. 1997 Apr;4(4):264-9. doi: 10.1016/s1076-6332(97)80027-x.

Abstract

RATIONALE AND OBJECTIVES

The authors assessed the reliability and reproducibility of volumetric measurements of olfactory bulbs and tracts (OBTs) and temporal lobes (TLs) with magnetic resonance (MR) imaging.

METHODS

Repetitive MR imaging of two phantoms and OBTs and TLs of patients was performed. Regions of interest were manually drawn around the relevant structures, and their volumes were measured at workstations. Intra- and interobserver variability for the two readers were measured. Intraclass and Pearson correlation coefficients and mean percentage differences were calculated.

RESULTS

The measured phantom volumes were within 1.9%-12% of the true volumes with a variability of < or = 5%. Intraclass and Pearson correlation coefficients were > or = 0.919 for measurements by a single reader and > or = 0.924 for measurements by different readers. The inter- and intraobserver variabilities were < or = 4.2% for TL and 11.3%-14.6% for OBTs.

CONCLUSION

Volumetric measurements of the olfactory apparatus can be reliably and accurately reproduced from MR images.