Izumi A, Minakami H, Sato I
Department of Obstetrics and Gynecology, Jichi Medical School, Tochigi, Japan.
J Clin Monit. 1997 Mar;13(2):103-8. doi: 10.1023/a:1007370920986.
To evaluate the accuracy of a new pulse oximeter in estimating the oxyhemoglobin (O2Hb) concentration in fetal arterial blood (SaO2) with either a high concentration of hemoglobin F (HbF) or a low concentration of O2Hb during labor.
Fetal forehead arterial oxygenation was determined in 44 fetuses by reflectance pulse oximeter (SpO2) during labor and was compared with SaO2 determined immediately after birth. Because HbF has little or no known effect on pulse oximetry, but does affect the laboratory multiwavelength "CO-oximeter" type reading, the SaO2 was corrected by HbF concentration. SpO2 and SaO2 were simultaneously measured in five hypoxic adult volunteers achieved by inhaling 11% oxygen.
A gradual decline in HbF concentration was seen during weeks 37 to 40 of gestation. HbF concentration varied near term, ranging between 53 and 88% of the fetal hemoglobin concentration (mean +/- SD = 74.6 +/- 6.3%). This alteration produced a lower %O2Hb percent by 4% at the most. The corrected SaO2 in cord blood correlated with fetal SpO2 (y = 0.974, x -7.279, r = 0.90). In five adults, SpO2 reflected well SaO2 with a mean +/- SD of bias of -1.1 +/- 2.9%.
SpO2 determined by a new reflectance pulse oximeter at the end of labor correlated with an immediate post-natal cord arterial blood sample before the first breath, with a mean and SD of bias of 8.5 +/- 6.2%. Reflectance pulse oximetry is a useful tool for continuous noninvasive monitoring of the fetal oxygen status during labor.
评估一种新型脉搏血氧仪在分娩过程中,对于血红蛋白F(HbF)浓度高或氧合血红蛋白(O2Hb)浓度低的胎儿动脉血中氧合血红蛋白(O2Hb)浓度(SaO2)估计的准确性。
在44例胎儿分娩过程中,通过反射式脉搏血氧仪(SpO2)测定胎儿前额动脉氧合情况,并与出生后立即测定的SaO2进行比较。由于HbF对脉搏血氧测定几乎没有已知影响,但会影响实验室多波长“CO血氧仪”类型的读数,因此根据HbF浓度对SaO2进行校正。对5名通过吸入11%氧气造成低氧状态的成年志愿者同时测量SpO2和SaO2。
妊娠37至40周期间可见HbF浓度逐渐下降。HbF浓度在足月时有所变化,范围为胎儿血红蛋白浓度的53%至88%(平均±标准差 = 74.6±6.3%)。这种变化最多使O2Hb百分比降低4%。脐血中校正后的SaO2与胎儿SpO2相关(y = 0.974,x - 7.279,r = 0.90)。在5名成年人中,SpO2能很好地反映SaO2,偏差的平均±标准差为 - 1.1±2.9%。
分娩结束时通过新型反射式脉搏血氧仪测定的SpO2与出生后第一声呼吸前立即采集的脐动脉血样本相关,偏差的平均值和标准差为8.5±6.2%。反射式脉搏血氧测定法是分娩期间连续无创监测胎儿氧状态的有用工具。