Plummer J L, Ilsley A H, Fronsko R R, Owen H
Department of Anaesthesia and Intensive Care, Flinders Medical Centre, Bedford Park, South Australia, Australia.
J Clin Monit. 1997 Mar;13(2):109-13. doi: 10.1023/a:1007303104811.
The Nellcor N-3000 pulse oximeter is designed to be able to identify signal artefact related to movement of the body part to which the probe is attached. It may therefore provide a reliable means of monitoring arterial oxyhemoglobin saturation (SpO2) in awake, moving patients. This study compared the Nellcor N-3000 and N-200 pulse oximeters in terms of their ability to identify readings associated with movement, in a group of volunteers making standardized movements.
Thirty-six volunteers were studied. Volunteers breathed room air throughout the study. SpO2 of each volunteer was monitored by both a Nellcor N-200 and a Nellcor N-3000 simultaneously on both hands. Volunteers made a series of five standardized movements, each lasting one minute, with each hand during the monitoring session, while SpO2 and oximeter status were recorded from all four oximeters. The mean SpO2 reading was calculated during each movement. SpO2 readings which the oximeter identified as being associated with movement, pulse search not locked, sensor not attached, or break in communications were excluded from analysis.
The N-3000 rejected from 17 to 78% of readings taken during movement, compared to 0 to 2% with the N-200. Although the remaining readings of both types of oximeters were subject to some movement artefact, which led to spuriously low SpO2, this was significantly less with the N-3000.
The Nellcor N-3000 pulse oximeter is able, to some extent, to identify movement artefact. It should offer an advantage over the N-200 when monitoring moving patients.
Nellcor N - 3000脉搏血氧仪旨在能够识别与连接探头的身体部位运动相关的信号伪影。因此,它可能为监测清醒、活动患者的动脉血氧血红蛋白饱和度(SpO2)提供一种可靠的方法。本研究在一组进行标准化运动的志愿者中,比较了Nellcor N - 3000和N - 200脉搏血氧仪识别与运动相关读数的能力。
对36名志愿者进行了研究。在整个研究过程中,志愿者呼吸室内空气。用Nellcor N - 200和Nellcor N - 3000同时对每位志愿者的双手进行SpO2监测。在监测期间,志愿者用每只手进行一系列五个标准化动作,每个动作持续一分钟,同时记录所有四个血氧仪的SpO2和血氧仪状态。计算每个动作期间的平均SpO2读数。将血氧仪识别为与运动、脉搏搜索未锁定、传感器未连接或通信中断相关的SpO2读数排除在分析之外。
N - 3000排除了运动期间所取读数的17%至78%,而N - 200排除的比例为0%至2%。尽管两种类型的血氧仪的其余读数都受到一些运动伪影的影响,导致SpO2假性降低,但N - 3000的这种情况明显较少。
Nellcor N - 3000脉搏血氧仪在一定程度上能够识别运动伪影。在监测活动患者时,它应比N - 200具有优势。