Suppr超能文献

使用一种考虑了运动伪影的新一代脉搏血氧仪监测健康婴儿的血红蛋白氧饱和度。

Monitoring of haemoglobin oxygen saturation in healthy infants using a new generation pulse oximeter which takes motion artifacts into account.

作者信息

Meyts Isabelle, Reempts Patrick Van, Boeck Kris De

机构信息

Department of Paediatric Pulmonology, University Hospital Gasthuisberg, Leuven, Herestraat 49, 3000 Leuven, Belgium.

出版信息

Eur J Pediatr. 2002 Dec;161(12):653-5. doi: 10.1007/s00431-002-1089-3. Epub 2002 Nov 6.

Abstract

UNLABELLED

The aim of this study was to establish normal values for overnight oxygen saturation (SpO2) in healthy term infants using an oximeter which takes into account motion artifacts and to compare these to normal values collected with a previous generation oximeter not correcting for motion artifacts. We recorded overnight SpO2 in 26 term, healthy infants (median age 136 days, range 6-364 days) in the home environment using the Nellcor Symphony N 3000 pulse oximeter with an averaging time of 3 s. A sample rate of 5 s was chosen. Motion artifacts were excluded from the analysis. Data were compared with those from a previous study, using the same inclusion and exclusion criteria with the Oxford Medilog. Median (quartiles) SpO2 was 98% (97%-99%). Median percentage of study time below SpO2 94% was 0.2% (0.1%-0.7%); median percentage of study time below SpO2 90% was 0.0% (0.0%-0.01%). Median SpO2with the Oxford oximeter was 97% (96%-98%); percentage of study time below SpO2 94% was 8% (2%-14%); percentage of study time below SpO2 90% was 2% (0%-4%). These data were compared with the Nellcor Symphony data: differences in median SpO2 were significant ( P<0.05); differences in percentage of time below SpO2 94% and 90% were also statistically significant ( P<0.001).

CONCLUSION

we established normal values of oxygen saturation in healthy term infants using the Nellcor Symphony 3000 pulse oximeter. Care should be taken in interpreting values obtained with different types of pulse oximeters.

摘要

未标注

本研究的目的是使用一种能考虑运动伪影的血氧仪,确定健康足月儿夜间血氧饱和度(SpO2)的正常值,并将其与使用上一代未校正运动伪影的血氧仪收集的正常值进行比较。我们在家中环境下,使用Nellcor Symphony N 3000脉搏血氧仪,对26名足月儿、健康婴儿(中位年龄136天,范围6 - 364天)进行了夜间SpO2记录,平均时间为3秒。选择了5秒的采样率。分析中排除了运动伪影。数据与之前一项研究的数据进行了比较,该研究使用Oxford Medilog时采用了相同的纳入和排除标准。SpO2的中位数(四分位数)为98%(97% - 99%)。SpO2低于94%的研究时间中位数为0.2%(0.1% - 0.7%);SpO2低于90%的研究时间中位数为0.0%(0.0% - 0.01%)。使用Oxford血氧仪时SpO2的中位数为97%(96% - 98%);SpO2低于94%的研究时间百分比为8%(2% - 14%);SpO2低于90%的研究时间百分比为2%(0% - 4%)。这些数据与Nellcor Symphony数据进行了比较:SpO2中位数的差异具有显著性(P<0.05);SpO2低于94%和90%的时间百分比差异也具有统计学显著性(P<0.001)。

结论

我们使用Nellcor Symphony 3000脉搏血氧仪确定了健康足月儿血氧饱和度的正常值。在解释使用不同类型脉搏血氧仪获得的值时应谨慎。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验