Experimental correlation-based identification of X-ray CT point spread function. Part 2: Simulation and design of input signal.

作者信息

Doré S, Kearney R E, De Guise J A

机构信息

Ecole de technologie supérieure, Montréal, Québec, Canada. sdore@mec. etsmtl.ca

出版信息

Med Biol Eng Comput. 1997 Jan;35(1):9-16. doi: 10.1007/BF02510385.

Abstract

The preferred signals for non-parametric correlation-based point spread function identification are white noise or pseudo-random binary sequences (PRBSs). Given the difficulty of building a phantom based on either of these signals, a new input is devised that corresponds to pseudo-randomly located holes. The positions of the holes correspond to zeros in a 2-D PRBS. To optimise the design of the phantom and to ensure proper imaging procedure, a number of simulations are conducted. The effects of the following parameters on identification quality are investigated: the size of the holes and their minimum separation, the period of the PRBS, input-output translational and rotational mis-registration, pixel size and the presence of cupping. The factors affecting identification quality the most are rotational alignment, hole size and separation, as well as sequence length. During simulations, a point spread function offering characteristics similar to the Philips Tomoscan CX is identified. Optimal results are obtained when the signal consists of 0.6 mm holes, separated by 0.9 mm, whose position is based on a 32 x 32 PRBS generated with a ten-stage shift-register. When adequate rotational alignment is provided, it is shown that the pseudo-randomly located holes signal is a good substitute for a purely white signal when identifying the PSF of a CT scanner.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验