Faber D B
J Chromatogr. 1977 Nov 11;142:421-30. doi: 10.1016/s0021-9673(01)92055-2.
The programmed multiple development thin-layer chromatography (PMD-TLC) technique was compared with the conventional chromatographic technique on both normal and HPTLC plates. The potency of this high-performance TLC is illustrated and discussed by means of the data obtained with the determination of digitoxin in human serum. Improvement of the efficiency of TLC, resulting in a better resolution and sensitivity with PMD-TLC using using HPTLC plates, makes this high-performance TLC technique comparable to high-performance liquid chromatography. This TLC approach might lead to a sensitive, rapid, selective and simply assay for routine serial analyses and, because of its specificity and flexibility, it may facilitate drug interaction studies.
在普通板和高效薄层色谱(HPTLC)板上,将程序控制多重展开薄层色谱(PMD-TLC)技术与传统色谱技术进行了比较。通过测定人血清中洋地黄毒苷所获得的数据,对这种高效薄层色谱的效能进行了说明和讨论。使用HPTLC板的PMD-TLC提高了薄层色谱的效率,带来了更好的分离度和灵敏度,使得这种高效薄层色谱技术可与高效液相色谱相媲美。这种薄层色谱方法可能会带来一种灵敏、快速、选择性强且简便的常规系列分析检测方法,并且由于其特异性和灵活性,可能会促进药物相互作用研究。