Blakeborough A, Ward J, Wilson D, Griffiths M, Kajiya Y, Guthrie J A, Robinson P J
Department of Radiology, St James's University Hospital, Leeds, England.
Radiology. 1997 Jun;203(3):759-65. doi: 10.1148/radiology.203.3.9169701.
To compare results with the following magnetic resonance (MR) imaging sequences in the detection of focal hepatic lesions: fast spin-echo (SE) before and after administration of superparamagnetic iron oxide (SPIO) particles, fat-suppressed T2-weighted SE, and dynamic gadolinium-enhanced fast low-angle shot (FLASH).
In 26 patients with known malignancy and documented focal liver lesions, 1.0-T MR imaging was performed prior to hepatic resection. All images were reviewed independently by four blinded observers. Sensitivity was calculated for each sequence and for each observer by means of alternative-free-response receiver operating characteristic (ROC) methods.
The mean area under the alternative-free-response ROC curve with SPIO-enhanced fast SE (0.85) was significantly greater (P < .02) than that with all other sequences. Detection was significantly improved with SPIO-enhanced fast SE compared with dynamic gadolinium-enhanced FLASH (which was the best of the other three sequences) for all lesions (P < .002) and for malignant lesions (P < .0002).
Findings with the SPIO-enhanced fast SE sequence improved detection of focal liver lesions and had the highest diagnostic accuracy.
比较以下磁共振(MR)成像序列在检测肝脏局灶性病变中的结果:超顺磁性氧化铁(SPIO)颗粒给药前后的快速自旋回波(SE)、脂肪抑制T2加权SE以及动态钆增强快速低角度激发(FLASH)。
对26例已知患有恶性肿瘤且有肝脏局灶性病变记录的患者,在肝切除术前进行1.0-T MR成像检查。所有图像由四名不知情的观察者独立进行评估。通过无替代自由反应接收器操作特性(ROC)方法计算每个序列以及每位观察者的敏感度。
SPIO增强快速SE序列的无替代自由反应ROC曲线下平均面积(0.85)显著大于(P <.02)所有其他序列。与动态钆增强FLASH(其他三个序列中表现最佳的)相比,SPIO增强快速SE对所有病变(P <.002)和恶性病变(P <.0002)的检测均有显著改善。
SPIO增强快速SE序列的结果提高了肝脏局灶性病变的检测率,且具有最高的诊断准确性。