Evaluation of an on-line patient exposure meter in neuroradiology.

作者信息

Gkanatsios N A, Huda W, Peters K R, Freeman J A

机构信息

Department of Radiology, University of Florida, Gainesville 32610-0374, USA.

出版信息

Radiology. 1997 Jun;203(3):837-42. doi: 10.1148/radiology.203.3.9169713.

Abstract

PURPOSE

To assess the clinical performance and usefulness of an on-line patient exposure meter installed on a neuroradiologic biplane imaging system.

MATERIALS AND METHODS

A commercial on-line patient exposure meter was installed on each plane of a biplane neuroradiologic imaging system. The meter computed skin exposures on the basis of selected technique factors (tube potential and current) and information about patient location relative to the x-ray tube. Simulations were performed to measure the system accuracy with an angiographic anthropomorphic head phantom with the skin exposures measured with an ionization chamber. Skin doses were computed for 114 consecutive patients who underwent diagnostic and interventional neuroradiologic procedures.

RESULTS

Agreement between measured and computed skin exposures in fluoroscopy, plain radiography, and digital imaging was generally within 5% of the true skin dose. For all fluoroscopic and radiographic procedures, total median skin doses were 1.20 and 0.64 Gy for the frontal and lateral planes, respectively. In both planes, patient skin doses resulted primarily from digital subtraction angiographic acquisitions. In 29 (25%) patients, the skin dose exceeded 2.00 Gy, but no radiation-induced deterministic effects were observed.

CONCLUSION

An on-line patient exposure meter can provide accurate radiation skin dose data in patients undergoing diagnostic and therapeutic neuroradiologic procedures.

摘要

目的

评估安装在神经放射双平面成像系统上的在线患者剂量仪的临床性能和实用性。

材料与方法

在神经放射双平面成像系统的每个平面上安装一台商用在线患者剂量仪。该剂量仪根据选定的技术参数(管电压和电流)以及患者相对于X射线管的位置信息计算皮肤剂量。使用血管造影人体头部模型进行模拟,并用电离室测量皮肤剂量,以评估系统准确性。计算了114例接受诊断性和介入性神经放射检查的连续患者的皮肤剂量。

结果

透视、平片摄影和数字成像中测量的皮肤剂量与计算的皮肤剂量之间的一致性通常在真实皮肤剂量的5%以内。对于所有透视和摄影程序,额面和侧面的总中位皮肤剂量分别为1.20和0.64 Gy。在两个平面上,患者皮肤剂量主要来自数字减影血管造影采集。29例(25%)患者的皮肤剂量超过2.00 Gy,但未观察到辐射诱发的确定性效应。

结论

在线患者剂量仪可为接受诊断性和治疗性神经放射检查的患者提供准确的辐射皮肤剂量数据。