Suppr超能文献

[使用球囊填塞时鼻黏膜的微循环]

[Microcirculation of the nasal mucosa during use of balloon tamponade].

作者信息

Klinger M, Siegert R

机构信息

Klinik für HNO-Heilkunde der Med. Universität Lübeck.

出版信息

Laryngorhinootologie. 1997 Mar;76(3):127-30. doi: 10.1055/s-2007-997400.

Abstract

BACKGROUND

Nasal packings are commonly accepted in the treatment of severe epistaxis. Cuffed catheters are known to cause damage to the nasal mucosa most likely by interfering with tissue perfusion. In this study the effect of different pressure levels on local perfusion of septal mucosa is investigated.

METHOD

In 15 healthy subjects the blood flow in septal mucosa was measured by laser doppler flowmetry by positioning a cuffed epistaxis catheter into the nasal cavity with a laser probe attached to it. Increasing pressure was administered by injecting saline solution while continuously recording intraluminal pressure, perfusion, and filling volume. The local pressure affecting the septal mucosa at the moment of stalling perfusion was determined by subtracting the extranasal cuff pressure from the current intranasal cuff pressure at same inflation volumes.

RESULTS

Microcirculation of the septal mucosa stopped when the local pressure exceeded a value of Pmean = 42 mmHg. Individual variations (n = 15) were small (s = 9 mmHg). The intraluminal cuff pressure was measured to be about ten times higher due to the retraction force of the cuff. Spontaneous oscillations of the blood flow were reduced with increasing pressure to the blood vessels. Filling volumes up to 3.2 ml were sufficient to stop perfusion.

CONCLUSIONS

Cuffed nasal packings stop the blood flow in nasal mucosa even at low local pressures. Depending on the material characteristics of different cuffs the pressure to dilate the cuff may, however, be several times higher than the actual local pressure. This effect may cause problems in the proper use of cuffed catheters. Laser doppler flowmetry proved to be helpful in determining reproducible perfusion values.

摘要

背景

鼻填塞常用于治疗严重鼻出血。已知带套囊导管很可能通过干扰组织灌注而对鼻黏膜造成损伤。本研究调查了不同压力水平对鼻中隔黏膜局部灌注的影响。

方法

在15名健康受试者中,通过将带套囊的鼻出血导管置入鼻腔,并在其上连接激光探头,用激光多普勒血流仪测量鼻中隔黏膜的血流。通过注入生理盐水增加压力,同时持续记录管腔内压力、灌注和充盈量。通过在相同充气量下用当前鼻内套囊压力减去鼻外套囊压力,确定灌注停止时影响鼻中隔黏膜的局部压力。

结果

当局部压力超过Pmean = 42 mmHg时,鼻中隔黏膜的微循环停止。个体差异(n = 15)较小(s = 9 mmHg)。由于套囊的回缩力,管腔内套囊压力测量值约高十倍。随着血管压力增加,血流的自发振荡减少。高达3.2 ml的充盈量足以停止灌注。

结论

带套囊的鼻填塞即使在低局部压力下也会使鼻黏膜血流停止。然而,根据不同套囊的材料特性,扩张套囊所需的压力可能比实际局部压力高几倍。这种效应可能会在带套囊导管的正确使用中引起问题。激光多普勒血流仪被证明有助于确定可重复的灌注值。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验