Kraus M, Buckingham T A, Candinas R
Department of Internal Medicine, University Hospital, Zurich, Switzerland.
Pacing Clin Electrophysiol. 1997 Jun;20(6):1691-7. doi: 10.1111/j.1540-8159.1997.tb03541.x.
We noted a series of 12 consecutive patients with a DDD Genisis pacemaker that showed an unexpected and a relatively rapid fall in battery voltage and output as these devices approached end-of-life (EOL). Twenty-one of 24 leads were Vitatron Helifix leads and there was a relatively high mean threshold (atrial 2.5 +/- 0.94 V; ventricular 2.9 +/- 0.65 V). These devices were replaced after 65 +/- 12 months. During the 9.3 +/- 3.5 months before replacement, a striking fall in voltage from 2.7 +/- 0.04 V to 2.49 +/- 0.05 V was seen. Battery impedance rose from 3 +/- 1.2 K omega to 10.2 +/- 4.3 K omega during this same period. We unexpectedly observed a marked difference between programmed and telemetered output for both atrial (50%) and ventricular leads (30%). A discrepancy between measured and telemetered magnet rate was also seen. Despite this relatively rapid fall in battery voltage, several of these devices did not meet the manufacturer's recommended replacement time (RRT) criteria by magnet rate or according to the projected RRT determined by the relationship of battery impedance to current drain. These data have implications for the selection of RRT and EOL criteria for this device. Magnet rate measured by surface ECG was the safest indicator for RRT. Follow-up for this pulse generator should be increased to every 2 months when battery impedance is > 2 KOhms or if there is a difference between programmed and measured output amplitude of more than 15%. The data also highlight the effect of combining high threshold leads with modern pacemakers with relatively "small" batteries as well as certain problems with telemetered data.
我们记录了连续12例植入DDD Genesis起搏器的患者,这些患者的起搏器在接近使用寿命结束(EOL)时,电池电压和输出出现了意外且相对快速的下降。24根导线中有21根是Vitatron Helifix导线,平均阈值相对较高(心房2.5±0.94V;心室2.9±0.65V)。这些设备在65±12个月后进行了更换。在更换前的9.3±3.5个月期间,电压从2.7±0.04V显著下降至2.49±0.05V。同期电池阻抗从3±1.2KΩ升至10.2±4.3KΩ。我们意外地观察到,心房(50%)和心室导线(30%)的程控输出和遥测输出之间存在显著差异。测量的和遥测的磁铁频率之间也存在差异。尽管电池电压下降相对较快,但其中一些设备并未达到制造商推荐的更换时间(RRT)标准,无论是根据磁铁频率还是根据电池阻抗与电流消耗的关系确定的预计RRT。这些数据对该设备的RRT和EOL标准的选择具有启示意义。通过表面心电图测量的磁铁频率是RRT最安全的指标。当电池阻抗>2KΩ或程控输出幅度与测量输出幅度之间的差异超过15%时,应将该脉冲发生器的随访增加至每2个月一次。这些数据还突出了高阈值导线与电池相对“小”的现代起搏器相结合的影响以及遥测数据的某些问题。