[Evaluation of the Cell-DYN 35))].

作者信息

Costa M J, Reis A B, Silva C, Fernandes I, Lopes S, Júnior E

机构信息

Laboratório de Hematologia, Hospital Santa Cruz, Carnaxide.

出版信息

Acta Med Port. 1996 Oct-Dec;9(10-12):309-18.

PMID:9254527
Abstract

The Cell-Dyn 3500 is an automated haematology analyzer which quantitatively measures and computes haematological quantities including a full "five-part" white cell differential. It measures 22 parameters for erythrocytes, white blood cells and platelets, also giving the respective histograms. Evaluation of the Cell-Dyn 3500 was performed according to the International Committee for Standardization in Haematology (ICSH) norms, for a period of 5 weeks. A total of 1,235 samples were studied by comparison with the Coulter MaxM. The five-part white cell differential and the flagging system were estimated and compared with the smear examination of 506 samples, by four clinical pathologists trained in cytology. Good correlation was obtained within, between batches, and day-to-day, for the following parameters: red blood cells (RBC), haemoglobin (HGB), mean cell volume (MCV), white blood cells (WBC) and platelets (PLT). The accuracy was estimated (RBC, HGB, VGM, WBC and PLT) each day with three different levels of titrated controls with good results. The linearity was established for RBC, HGB, WBC and PLT. The results obtained were good. Carry-over studies were performed according to the Broughton method for the same parameters and the results were also good. Stability studies for the automatic parameters including the differential white blood cell count showed that these parameters were stable at 4 degrees C for 48 hours. At room temperature the stability was reduced to 7 hours. Agreement was good between the Cell-Dyn 3500 and the Coulter MaxM, for the automatic haemocytometric values. The comparative studies between the five-part white cell differential of the haematologic analyzer and the manual differential showed excellent results for neutrophils and lymphocytes, very good for monocytes and eosinophils. For the flag estimation two criteria were established, one based on the clinical significance and the other based on the alarm detection described in the analyzer manual. The specificity was good for both criteria. In general the sensibility was better for the second criteria. The Cell-Dyn 3500 has thus shown to be a good haematology analyser which greatly reduces the morphological examination of smears.

摘要

相似文献

1
[Evaluation of the Cell-DYN 35))].
Acta Med Port. 1996 Oct-Dec;9(10-12):309-18.
2
Complete analytical and diagnostic performances of the Abbott Cell Dyn 3500.雅培Cell Dyn 3500的完整分析和诊断性能。
Panminerva Med. 1998 Jun;40(2):116-25.
3
Performance evaluation of Abbott CELL-DYN Ruby for routine use.雅培CELL-DYN Ruby常规使用的性能评估。
Int J Lab Hematol. 2008 Oct;30(5):400-7. doi: 10.1111/j.1751-553X.2007.00971.x.
4
[Standardization Committee for Haematology. External Quality Assurance Program for General Hematology. Evaluation of the 1994 results].[血液学标准化委员会。普通血液学外部质量保证计划。1994年结果评估]
Sangre (Barc). 1996 Apr;41(2):115-23.
5
Performance evaluation of the Sysmex XS-1000i automated haematology analyser.希森美康XS-1000i全自动血液分析仪的性能评估
Int J Lab Hematol. 2009 Oct;31(5):560-6. doi: 10.1111/j.1751-553X.2008.01081.x. Epub 2009 Jun 18.
6
Performance evaluation of the Sysmex pocH-100i automated hematology analyzer.Sysmex pocH-100i全自动血液分析仪的性能评估
Lab Hematol. 2005;11(2):107-17. doi: 10.1532/LH96.04068.
7
Performance evaluation of BC-3200 hematology analyzer in a university hospital.BC-3200血液分析仪在某大学医院的性能评估
Int J Lab Hematol. 2008 Jun;30(3):205-13. doi: 10.1111/j.1751-553X.2007.00950.x.
8
Evaluation of the Beckman Coulter AcT 5 diff AL hematology analyzer in a hospital setting.贝克曼库尔特AcT 5 diff AL血液分析仪在医院环境中的评估
Lab Hematol. 2003;9(3):167-74.
9
Aging stability of complete blood count and white blood cell differential parameters analyzed by Abbott CELL-DYN Sapphire hematology analyzer.采用雅培CELL-DYN Sapphire血液分析仪分析全血细胞计数及白细胞分类参数的老化稳定性。
Int J Lab Hematol. 2009 Feb;31(1):87-96. doi: 10.1111/j.1751-553X.2007.01009.x. Epub 2008 Jan 7.
10
[Analytical evaluation of an automated hematologic analyzer: Cell Dyn 3500].
Minerva Med. 1996 Mar;87(3):123-30.