Suppr超能文献

Quality framework for force plate testing.

作者信息

Fleming H E, Hall M G, Dolan M J, Paul J P

机构信息

University College Dublin, Eire.

出版信息

Proc Inst Mech Eng H. 1997;211(3):213-9. doi: 10.1243/0954411971534322.

Abstract

Following work done in C.E.C. project AIM 2002 CAMARC II, a framework of the relevant standards for testing equipment used in laboratories analysing human movement is presented. In particular, gait analysis laboratories may provide a service for a clinical team considering treatment such as surgical intervention or provision of orthoses. In this case the analysis and equipment used in the analysis must have assured quality so that the correct clinical interpretation of the results is possible. The operation of a gait analysis laboratory may be described in quality standards terminology as the operation of a 'special process'. The operation of a special process requires that particular attention be paid to procedures including confirmation of the equipment and regular testing to ensure correct operation of the equipment. In general this is difficult to achieve since equipment manufacturers still have to address this technical requirement. Although difficult, these procedures must be accomplished in the operation of a special process. Passages of the most pertinent standards are presented and a methodology of a periodic confirmation and regular spot checking of equipment suggested to ensure its correct operation. Although force plates are specifically addressed, the framework could be applied to other dynamometers and measurement equipment.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验