Suppr超能文献

Sysmex NE - 8000与SF - 3000之间红细胞分布宽度中平均红细胞体积依赖性偏差。

Mean corpuscular volume-dependent deviations in red blood cell distribution width between Sysmex NE-8000 and SF-3000.

作者信息

Bartels P C, Schoorl M

机构信息

Department of Clinical Chemistry, Haematology and Immunology, Medical Centre Alkmaar, The Netherlands.

出版信息

Eur J Clin Chem Clin Biochem. 1997 Jul;35(7):549-50.

PMID:9263733
Abstract

Values for red blood cell distribution width yield information concerning size heterogeneity of red blood cell populations. Comparative studies between various haematology analysers are obligatory to quantify apparatus-dependent deviations with respect to accuracy and subsequent consequences with respect to clinical interpretation of results. In this study a comparative assessment is performed between a Sysmex SF-3000 and NE-8000 Haematology Analyser. Fifty-four specimens with abnormal red blood cell volumes and a hundred specimens from apparently healthy subjects were assayed. Systematic deviations in values concerning red blood cell distribution width are demonstrated to show a tendency to decrease towards higher mean corpuscular volume values.

摘要

红细胞分布宽度值可提供有关红细胞群体大小异质性的信息。必须对各种血液学分析仪进行比较研究,以量化仪器在准确性方面的依赖性偏差以及对结果进行临床解释的后续影响。在本研究中,对Sysmex SF - 3000和NE - 8000血液学分析仪进行了比较评估。对54份红细胞体积异常的标本和100份明显健康受试者的标本进行了检测。结果表明,红细胞分布宽度值的系统偏差呈现出随着平均红细胞体积值升高而降低的趋势。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验