Teslow T N
Department of Radiology, Baylor College of Medicine, Houston, TX 77030, USA.
J Digit Imaging. 1997 Aug;10(3 Suppl 1):128-32. doi: 10.1007/BF03168675.
This work evaluates differences in density linearity, contrast, and resolution of digital images acquired from a laser film scanner. A calibrated photographic step tablet tested linearity of diffuse optical density (OD). It was digitized within center openings of a dark and light film. The 14- x 17-inch films were scanned into 1024 x 1280 picture elements of 8 bits. A photographic tapered resolution guide evaluated the scanner's detail resolution. A single characteristic curve adequately represents the rate of change of OD per pixel value, but two piecewise linear functions are better. One ranges in light OD and a second in dark OD. Contrast shows a steeper slope in a light film rather than a dark film, yet a greater difference exists between two identical digitizers. The resolution test diminishes at 1.3 line pairs per mm (LP/mm), which compares to a Nyquist limit of 1.44 LP/mm. Such density, contrast, and resolution tests yield a benchmark to assure quality operations.
这项工作评估了从激光胶片扫描仪获取的数字图像在密度线性、对比度和分辨率方面的差异。使用经过校准的摄影阶梯平板测试漫射光密度(OD)的线性。它在暗片和明片的中心开口内进行数字化处理。将14×17英寸的胶片扫描成1024×1280像素的8位图像。使用摄影渐变分辨率指南评估扫描仪的细节分辨率。一条单一的特性曲线足以表示每像素值的OD变化率,但两段分段线性函数效果更好。一段用于浅OD范围,另一段用于深OD范围。对比度在明片中显示出比暗片中更陡的斜率,但在两台相同的数字化仪之间存在更大的差异。分辨率测试在每毫米1.3线对(LP/mm)时降低,而奈奎斯特极限为1.44 LP/mm。这种密度、对比度和分辨率测试产生了一个基准,以确保操作质量。