Halpern E J
Department of Radiology, Jefferson Medical College, Thomas Jefferson University, Philadelphia, PA, USA.
J Digit Imaging. 1995 Feb;8(1):3-9. doi: 10.1007/BF03168050.
Although clinical images provide the ultimate test of diagnostic performance for a film digitizer, such images are not generally suitable for daily quality control (QC) purposes. However, a well-designed test pattern will provide a rapid, comprehensive, objective and reproducible assessment of image quality. This pattern should evaluate various parameters of image quality, including high contrast resolution, low contrast discrimination, linearity of gray scale, geometric distortion, and noise. Furthermore, the pattern should detect light leak and film slippage, two problems commonly associated with film digitizers. The test pattern described in this manuscript was designed to provide quantitative measures of performance for a film digitizer. As part of a regular QC routine for a laser scanner or charge-coupled device digitizer, this pattern provides a simple method to identify and quantify changes in digital image quality.
尽管临床图像为胶片数字化仪的诊断性能提供了最终检验,但此类图像通常不适用于日常质量控制(QC)目的。然而,精心设计的测试图案将对图像质量提供快速、全面、客观且可重复的评估。该图案应评估图像质量的各种参数,包括高对比度分辨率、低对比度辨别力、灰度线性、几何失真和噪声。此外,该图案应能检测漏光和胶片滑动这两个与胶片数字化仪常见相关的问题。本手稿中描述的测试图案旨在为胶片数字化仪的性能提供定量测量。作为激光扫描仪或电荷耦合器件数字化仪常规QC程序的一部分,该图案提供了一种识别和量化数字图像质量变化的简单方法。