Halpern E J, Esser P D
Department of Radiology, Columbia Presbyterian Medical Center, New York, NY.
J Digit Imaging. 1991 Nov;4(4):241-7. doi: 10.1007/BF03173906.
Various quality control (QC) procedures may be used to evaluate image quality for picture archival and communications (PACS) systems. A standard PACS QC protocol applied on a regular basis is desirable to ensure optimal diagnostic performance. We describe a QC phantom designed especially to test PACS systems that acquire images by digitizing x-ray films. The phantom is a sheet of x-ray film upon which a digital test pattern is printed. Multiple parameters of image quality are tested, including resolution, contrast, gray scale, geometric distortion, and noise. Individual test patterns are incorporated to detect specific artifacts of laser scanner digitizers. As part of a regular QC protocol, the phantom provides an objective measurement of change in digital image quality over time, as well as an objective means for comparison with other systems.
可采用各种质量控制(QC)程序来评估图像存档与通信系统(PACS)的图像质量。定期应用标准的PACS QC协议有助于确保最佳诊断性能。我们描述了一种专门设计用于测试通过数字化X线胶片获取图像的PACS系统的QC体模。该体模是一张印有数字测试图案的X线胶片。测试图像质量的多个参数,包括分辨率、对比度、灰度、几何失真和噪声。纳入了单独的测试图案以检测激光扫描仪数字化仪的特定伪影。作为常规QC协议的一部分,该体模可客观测量数字图像质量随时间的变化,以及与其他系统进行比较的客观方法。