Man K C, Daoud E G, Knight B P, Bahu M, Weiss R, Zivin A, Souza S J, Goyal R, Strickberger S A, Morady F
Department of Internal Medicine, University of Michigan Medical Center, Ann Arbor 48109-0022, USA.
J Cardiovasc Electrophysiol. 1997 Sep;8(9):974-9. doi: 10.1111/j.1540-8167.1997.tb00619.x.
The purpose of this study was to determine the accuracy of the unipolar electrogram for identifying the earliest site of ventricular activation. The earliest site of ventricular activation may be identified with the unipolar electrogram by the absence of an R wave. However, the accuracy of this technique is unknown.
A single ventricular premature complex was induced mechanically at the tip of an electrode catheter to simulate a ventricular premature depolarization site of origin. Unipolar electrograms were recorded from the right ventricular septum at the tip electrode and at 2, 5, 8, and 11 mm from the electrode tip in 20 patients. No R waves were detected at the ventricular premature depolarization site of origin. R waves were detected in 4 of 20 patients (20%) at 2 mm from the tip electrode and 7 of 20 patients (35%) at 5, 8, and 11 mm from the tip electrode. An R wave was not observed at distances < or = 11 mm from the site of tachycardia origin in 13 of 20 patients (65%).
While an R wave in the unipolar electrogram can be seen as close as 2 mm from the site of impulse origin, the absence of an R wave as an indicator of the site of impulse origin in the right ventricle is highly inaccurate. Therefore, the absence of an R wave in the unipolar electrogram is unlikely to be an adequate guide for identification of an effective target site for ablation of right ventricular tachycardia.
本研究的目的是确定单极电图用于识别心室激动最早部位的准确性。单极电图可通过无R波来识别心室激动的最早部位。然而,该技术的准确性尚不清楚。
在20例患者中,通过机械刺激电极导管尖端诱发单个室性早搏复合波,以模拟室性早搏的起源部位。从右心室间隔的尖端电极以及距电极尖端2、5、8和11毫米处记录单极电图。在室性早搏起源部位未检测到R波。在距尖端电极2毫米处,20例患者中有4例(20%)检测到R波;在距尖端电极5、8和11毫米处,20例患者中有7例(35%)检测到R波。在20例患者中的13例(65%),在距心动过速起源部位≤11毫米处未观察到R波。
虽然在距冲动起源部位近至2毫米处可在单极电图中看到R波,但在右心室中,将无R波作为冲动起源部位的指标准确性极低。因此,单极电图中无R波不太可能成为识别右心室心动过速有效消融靶点的充分指导。