Marks L S, Dorey F J, Macairan M L, Park C, deKernion J B
Urological Sciences Research Foundation, Culver City, CA 90232, USA.
Urology. 1997 Sep;50(3):341-8. doi: 10.1016/S0090-4295(97)00293-8.
We sought to assess the accuracy, reliability, and clinical utility of the noninvasive determination of bladder volume using an automated, compact three-dimensional (3-D) ultrasound device.
We prospectively tested 249 adult outpatients for accuracy (n = 182), by comparing scan versus catheter volumes, or reliability (n = 67), by comparing the scan readings of two independent observers. Two models of the bladder scan device were tested (BVI-2500, 1994 and 1995 models).
Scan and catheter volumes were correlated (y = 1.02x + 12.6, R2 = 0.90, P < 0.001) across a range of zero to 1015 cc, regardless of which machine model was used. Scan volume underestimated catheter volume by an average of 10 cc in men and 20 cc in women. If a scan-predicted volume of 100 cc or greater were used as a cutpoint for clinical importance, the device exhibited a sensitivity of 97%, a specificity of 91%, and an overall accuracy of 94%. These results were not affected by age, gender, height, weight, diagnosis, uterine presence/prostate size, or user experience. The two observers, one a graduate physician and the other a college student, achieved essentially the same volume determinations (y = 0.96x + 0.13, R2 = 0.90, P < 0.001).
Volume determinations obtained with the bladder scan device are accurate and reliable in adult outpatients. A special technician is not required. These results may be attributable to use of automated planimetry and 3-D volumetry, rather than a fixed geometric formula, to custom measure each bladder shape.
我们试图评估使用自动化、紧凑型三维(3-D)超声设备无创测定膀胱容量的准确性、可靠性和临床实用性。
我们前瞻性地测试了249名成年门诊患者的准确性(n = 182),通过比较扫描测量的膀胱容量与导尿管测量的膀胱容量;或测试可靠性(n = 67),通过比较两名独立观察者的扫描读数。对两种型号的膀胱扫描设备(BVI - 2500,1994年和1995年型号)进行了测试。
在0至1015 cc的范围内,无论使用哪种机器型号,扫描测量的膀胱容量与导尿管测量的膀胱容量都具有相关性(y = 1.02x + 12.6,R2 = 0.90,P < 0.001)。扫描测量的膀胱容量在男性中平均比导尿管测量的容量低10 cc,在女性中低20 cc。如果将扫描预测容量100 cc或更大作为具有临床意义的切点,则该设备的灵敏度为97%,特异性为91%,总体准确率为94%。这些结果不受年龄、性别、身高、体重、诊断、子宫存在情况/前列腺大小或使用者经验的影响。两名观察者,一名是住院医师,另一名是大学生,得出的容量测定结果基本相同(y = 0.96x + 0.13,R2 = 0.90,P < 0.001)。
使用膀胱扫描设备获得的容量测定结果在成年门诊患者中准确且可靠。不需要专门的技术人员。这些结果可能归因于使用自动平面测量法和三维容积测量法,而非固定的几何公式,来针对每个膀胱形状进行定制测量。