Suppr超能文献

热释光晶体LiF:Mg,Ti和LiF:Mg,Cu,P用于临床剂量测定的有效性比较。

Comparison of effectiveness of thermoluminescent crystals LiF:Mg,Ti, and LiF:Mg,Cu,P for clinical dosimetry.

作者信息

Harris C K, Elson H R, Lamba M A, Foster A E

机构信息

Department of Radiation Oncology, University of Cincinnati, Ohio 45267-0757, USA.

出版信息

Med Phys. 1997 Sep;24(9):1527-9. doi: 10.1118/1.598042.

Abstract

This study compared the relative effectiveness of TLD crystals LiF:Mg,Ti (TLD-100) and LiF:Mg,Cu,P (TLD-700H) for clinical dosimetry, focusing on reproducibility, linearity, and energy response. Experimental results indicated that TLD-700H was superior to TLD-100 with regard to reproducibility, lack of supralinearity, and the absence of variation in TL signal with radiation quality. TLD-700H also had the additional advantages of higher sensitivity and immediate readability. The investigators conclude that this relatively new TLD crystal shows promising potential for clinical dosimetry.

摘要

本研究比较了热释光剂量计(TLD)晶体LiF:Mg,Ti(TLD - 100)和LiF:Mg,Cu,P(TLD - 700H)在临床剂量测定中的相对有效性,重点关注再现性、线性和能量响应。实验结果表明,TLD - 700H在再现性、无超线性以及热释光(TL)信号随辐射质量无变化方面优于TLD - 100。TLD - 700H还具有更高灵敏度和即时可读性的额外优势。研究人员得出结论,这种相对较新的TLD晶体在临床剂量测定中显示出有前景的潜力。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验