Abstract

The Varian CadPlan algorithm for computation of relative dose distributions and monitor unit calculations for Enhanced Dynamic Wedge (EDW) fields is based on a combination of open field beam data and Segmented Treatment data Tables. Calculation of dose by the pencil beam convolution model uses scatter kernels and boundary kernels to create the distribution. The principles of the pencil beam convolution model is presented. Comparison of measured and calculated monitor units and relative dose distributions showed good agreement and the deviations are within international accepted tolerans. Test results indicate that the EDW model works satisfactorily for all energies and wedge angles.

摘要

用于增强动态楔形野(EDW)的相对剂量分布计算和监测单位计算的瓦里安CadPlan算法基于开放野束流数据和分段治疗数据表的组合。通过笔形束卷积模型计算剂量时使用散射核和边界核来创建分布。介绍了笔形束卷积模型的原理。测量和计算的监测单位及相对剂量分布的比较显示出良好的一致性,偏差在国际公认的容差范围内。测试结果表明,EDW模型在所有能量和楔形角度下均能令人满意地工作。