A phase I study of Femcap used with and without spermicide. Postcoital testing.

作者信息

Mauck C K, Baker J M, Barr S P, Johanson W, Archer D F

机构信息

Contraceptive Research and Development Program (CONRAD), Eastern Virginia Medical School, Arlington, USA.

出版信息

Contraception. 1997 Aug;56(2):111-5. doi: 10.1016/s0010-7824(97)00098-x.

Abstract

The objectives of the study were to assess the ability of the Femcap, a new vaginal contraceptive device made of silicone and designed to fit snugly around the cervix to prevent the penetration of sperm into midcycle cervical mucus when used with and without spermicide; and to compare it with the standard contraceptive diaphragm used with spermicide. Eight women underwent two baseline cycles of postcoital testing in which no device was used, followed by three test cycles in which Femcap with spermicide, Femcap with nonspermicidal lubricant (K'Y gel) or the Ortho All-Flex diaphragm with spermicide was used. The sequence of testing cycles was randomized. In each cycle, condoms were used prior to midcycle, then a midcycle cervical mucus specimen was examined to ensure midcycle characteristics and the absence of sperm. Each woman then had intercourse using either no device (baseline cycles) or the prescribed device (test cycles) and returned 2-3 h afterwards. Cervical mucus was again assessed for adequacy and the presence of spermatozoa. The average number of progressively motile sperm seen per high power field was as follows: first baseline cycle, 18.0; second baseline cycle, 17.8; test cycle with Femcap used with nonspermicidal lubricant, 0.1; test cycle with Femcap used with spermicide, 0.2; and test cycle with the diaphragm used with spermicide, 0.0. There was no significant difference between baseline cycles or among test cycles in the average number of progressively motile sperm seen (p > 0.05). The average number of progressively motile sperm seen in each test cycle did, however, differ significantly from the average number seen in either baseline cycle (p < 0.05). Femcap, used with either a spermicidal lubricant or a nonspermicidal lubricant, appears to be comparable with the diaphragm used with spermicide in preventing sperm from entering midcycle cervical mucus.

摘要

该研究的目的是评估Femcap(一种由硅胶制成、设计为紧密贴合宫颈的新型阴道避孕装置)在使用和不使用杀精剂时防止精子穿透排卵期宫颈黏液的能力;并将其与使用杀精剂的标准避孕隔膜进行比较。八名女性进行了两个未使用任何装置的性交后测试基线周期,随后进行三个测试周期,分别使用含杀精剂的Femcap、含非杀精润滑剂(K'Y凝胶)的Femcap或含杀精剂的奥豪全柔隔膜。测试周期的顺序是随机的。在每个周期中,排卵期前使用避孕套,然后检查一份排卵期宫颈黏液样本,以确保具有排卵期特征且无精子。然后,每位女性在不使用任何装置(基线周期)或使用指定装置(测试周期)的情况下进行性交,并在2至3小时后返回。再次评估宫颈黏液是否充足以及是否存在精子。每个高倍视野中观察到的活动精子平均数如下:第一个基线周期为18.0;第二个基线周期为17.8;使用含非杀精润滑剂的Femcap的测试周期为0.1;使用含杀精剂的Femcap的测试周期为0.2;使用含杀精剂的隔膜的测试周期为0.0。观察到的活动精子平均数在基线周期之间或测试周期之间没有显著差异(p>0.05)。然而,每个测试周期中观察到的活动精子平均数与任一基线周期中观察到的平均数有显著差异(p<0.05)。使用杀精润滑剂或非杀精润滑剂的Femcap在防止精子进入排卵期宫颈黏液方面似乎与使用杀精剂的隔膜相当。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验