可行性研究:使用锎252进行体内中子活化以进行钙的区域测量。

Feasibility study: in vivo neutron activation for regional measurement of calcium using californium 252.

作者信息

Evans H J, LeBlanc A D, Johnson P C

出版信息

Med Phys. 1976 May-Jun;3(3):148-52. doi: 10.1118/1.594218.

Abstract

The feasibility of using a collimated 252Cf neutron source to measure regional changes in skeletal calcium was tested because in vivo regional activation of diseased bone should offer advantages over the more widely reported total-body calcium measuring techniques. Regional activation allows examination of discrete regions where the greatest changes in calcium content occur. Additionally, a simpler radiation facility is required for regional studies. Using a 5.5-mug 252Cf source, thermal neutron flux and absorbed dose were measured in a tissue-equivalent phantom. Detection efficiency of 49Ca gamma rays for conditions simulating regional activation were measured using a 29-cm-diameter X 10-cm-thickness sodium iodide detector. These in vitro measurements indicate that a collimated 252Cf source can be used for regional neutron activation of the lower spine and legs. Preliminary calculations indicate that a 1-3-mg source provides adequate count rates for statistical accuracy with a bone marrow dosage acceptable for human patients and normal subjects.

摘要

由于患病骨骼的体内区域活化相对于更广泛报道的全身钙测量技术应具有优势,因此测试了使用准直的²⁵²Cf中子源测量骨骼钙区域变化的可行性。区域活化允许检查钙含量变化最大的离散区域。此外,区域研究需要更简单的辐射设施。使用5.5微克的²⁵²Cf源,在组织等效模型中测量热中子通量和吸收剂量。使用直径29厘米×厚度10厘米的碘化钠探测器测量模拟区域活化条件下⁴⁹Caγ射线的探测效率。这些体外测量表明,准直的²⁵²Cf源可用于下脊柱和腿部的区域中子活化。初步计算表明,1 - 3毫克的源可为统计准确性提供足够的计数率,且骨髓剂量对人类患者和正常受试者来说是可接受的。