Characteristics of mixed venous oxygen saturation and physical activity as parameters for artificial heart control.

作者信息

Ohno T, Masuzawa T, Nakamura M, Tatsumi E, Taenaka Y, Nakatani T, Wakisaka Y, Nishimura T, Takewa Y, Takano H, Fukui Y

机构信息

Department of Artificial Organs, National Cardiovascular Center Research Institute, Osaka, Japan.

出版信息

ASAIO J. 1997 Sep-Oct;43(5):M677-81.

PMID:9360132
Abstract

Mixed venous oxygen saturation (SvO2) and physical activity (PhyAc) are practical candidates as parameters of total artificial heart (TAH) control because SvO2 can be measured through a transparent blood pump housing with infrared rays and PhyAc can be calculated from signals of an accelerometer used for rate response pacemakers. Although the methods for measurement of the parameters have already been developed, characteristics of these parameters for TAH control, such as during exercise, are still unclear and were examined in this study. SvO2, cardiac output (CO), and PhyAc were measured as parameters. Multi-stage treadmill exercise tests were performed. Difference values (DVs) from the value at the start of exercise showed better correlation than did absolute values. Correlation coefficients between DV in CO and DV in SvO2 and between DV in CO and PhyAc were high at -0.82 and 0.72, while the time constants for the change of SvO2 and CO to the PhyAc change were 26 and 32 sec. Although the correlation coefficient between the CO and SvO2 was higher than that between CO and PhyAc, PhyAc responded more quickly to the speed change compared with the response of SvO2 and CO. It was concluded that SvO2 and PhyAc were useful parameters with different characteristics for TAH control during exercise.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验