PURPOSE

To determine the intraindividual and interindividual characteristics of normal sensitivity derived by short-wavelength automated perimetry (SWAP) as a function of threshold algorithm. To determine also the influence of ocular media absorption on the magnitude of the interindividual variation in normal sensitivity, and hence the confidence limits, derived by SWAP.

METHODS

The sample comprised 51 normal subjects, stratified for age by decade (mean age, 55.5 years; range, 24-83 years) and experienced in white-on-white (W-W) perimetry and SWAP. One randomly assigned eye of each subject was examined on three occasions with Program 30-2 of the 640 Humphrey Field Analyzer using the Full Threshold and FASTPAC strategies for SWAP and W-W perimetries. Ocular media absorption (OMA) was assessed by the difference in scotopic sensitivity to stimuli of 410 and 560 nm.

RESULTS

The group mean examination time (P < 0.001) was greater for SWAP than for W-W perimetry for both the Full Threshold (15.0% longer) and FASTPAC strategies (16.8% longer). The gradient of the age-decline in Mean Sensitivity for SWAP was approximately 25% less steep when corrected for OMA than when uncorrected. The interindividual normal variability, expressed as the coefficient of variation, for SWAP without correction for OMA was 2.7 times greater (range 2.0-3.9), and with correction 1.9 times greater (range 1.4-2.9), than that for W-W perimetry.

CONCLUSIONS

The increased interindividual normal variability of SWAP, exacerbated by the lack of correction for OMA, currently limits the utility of SWAP in that the reduction in sensitivity required to indicate abnormality was proportionately greater than for W-W perimetry.

摘要

目的

确定短波长自动视野计(SWAP)根据阈值算法得出的正常敏感度的个体内和个体间特征。同时确定眼介质吸收对SWAP得出的正常敏感度个体间差异大小的影响,进而确定其置信限。

方法

样本包括51名正常受试者,按年龄每十年分层(平均年龄55.5岁;范围24 - 83岁),且有白色视标(W-W)视野检查和SWAP检查经验。使用640型Humphrey视野分析仪的程序30-2,对每位受试者一只随机分配的眼睛进行三次检查,分别采用全阈值和FASTPAC策略进行SWAP和W-W视野检查。通过对410和560nm刺激的暗视敏感度差异评估眼介质吸收(OMA)。

结果

对于全阈值策略(长15.0%)和FASTPAC策略(长16.8%),SWAP的组平均检查时间均比W-W视野检查长(P < 0.001)。校正OMA后,SWAP平均敏感度随年龄下降的梯度比未校正时约平缓25%。未校正OMA时,SWAP的个体间正常变异性(以变异系数表示)比W-W视野检查大2.7倍(范围2.0 - 3.9),校正后大1.9倍(范围1.4 - 2.9)。

结论

SWAP个体间正常变异性增加,且因未校正OMA而加剧,目前限制了SWAP的实用性,因为指示异常所需的敏感度降低幅度相对于W-W视野检查成比例地更大。