Measurement of percentage depth dose and lateral beam profile for kilovoltage x-ray therapy beams.

作者信息

Li X A, Ma C M, Salhani D

机构信息

Department of Medical Physics, Ottawa Regional Cancer Centre, Canada.

出版信息

Phys Med Biol. 1997 Dec;42(12):2561-8. doi: 10.1088/0031-9155/42/12/019.

Abstract

In this work, nine commonly used dosimetry detectors have been investigated to determine suitable relative dosimeters for kilovoltage x-ray beams. By comparison with the Monte Carlo calculated data, it was determined that for the detectors studied the PTW N23342, Markus and NACP parallel-plate chambers are more suitable for the measurement of percentage depth dose (PDD) data for this beam quality range with an uncertainty of about 3%. A diode detector may be used to measure the PDD for the 100 kVp beam, but it is not suitable for higher energies (300 kVp). The Capintec parallel-plate chamber may be adequate for medium-energy photons, but it has a slightly higher uncertainty for low-energy x-rays (100 kVp). For the measurement of beam profiles, diode and film yield incorrect profile tails, which can be corrected using the RK ionization chamber.