Krapp H, Breuker K H
Geburtshilfe Frauenheilkd. 1976 Apr;36(4):344-7.
A case of fetal death in utero is discussed in which fetal heart tones were picked up with the cardiotocograph via direct leads from the scalp electrode during labor induction, i.e., during the course of labor. After delivery, however, it was apparent that the fetus had been dead for some time (third grade maceration). Technical errors originating in the instrument itself could be excluded since the instrument was used successfully for other deliveries before and after this particular birth. A defect caused by the scalp electrode was also not possible since the electrode was changed three times.
本文讨论了一例子宫内胎儿死亡的病例。在引产过程中,即分娩过程中,通过头皮电极的直接导联,用胎心监护仪检测到了胎儿心音。然而,分娩后发现胎儿已经死亡一段时间(Ⅲ度浸软)。由于该仪器在此次特殊分娩前后均成功用于其他分娩,因此可以排除仪器本身产生的技术误差。由于头皮电极更换了三次,因此也不可能是由头皮电极引起的缺陷。