Erhart P, Ladd M E, Steiner P, Heske N, Dumoulin C L, Debatin J F
University Hospital, Zürich, Switzerland.
Magn Reson Med. 1998 Feb;39(2):279-84. doi: 10.1002/mrm.1910390215.
An improved MR tracking coil design is described that provides more robust tracking performance. It is shown theoretically and experimentally that a coil equipped with an internal spin source increases the signal-to-noise ratio in comparison to a coil system without internal source. The tracking behavior is stable even in air, which is important for laparoscopic work, in which the device has to be placed inside the inflated abdomen. Two types of coil windings were tested: one with the windings perpendicular to the coil axis and the other one with the windings tilted relative to this axis. The experiments showed no significant effect on the signal-to-noise ratio between the two types. The improved MR-tracking coil design with internal source was successfully used in cholecystostomies and in laparoscopic interventions; both procedures were performed on swine.