Haslett R S, Batterbury M, Cuypers M, Cooper R L
University of Liverpool, UK.
Eye (Lond). 1997;11 ( Pt 5):692-7. doi: 10.1038/eye.1997.179.
To assess the inter-observer agreement of the measurement of optic disc dimensions by two observers using a modified 60 dioptre (D) fundus examination lens.
The vertical disc and cup diameters of 29 eyes were measured by two independent observers using a 60 D lens modified by incorporation of a 0.1 millimetre scale graticule. The vertical cup/disc ratio was calculated. Inter-observer agreement was assessed by calculation of the inter-observer differences and by the weighted kappa statistic.
The two observers showed good agreement for the measurement of disc diameter (mean difference -0.04; range -0.04, 0.2) and for cup diameter (mean difference -0.03; range 0.3, 0.2). Closer agreement for the vertical cup/disc ratio was achieved (mean kappa 0.96; 95% confidence limits 0.90, 1.0). The 95% confidence limit for the mean inter-observer difference in cup/disc ratio was 0.11, suggesting that a change of > 0.1 in the assessment of the cup/disc ratio by this technique is significant at the 5% level.
High inter-observer agreement of optic disc measurement can be achieved with this technique. The method has the potential to improve the clinical evaluation of the optic disc and the precision and accuracy of the clinical measurement of other fundal structures.
评估两名观察者使用改良的60屈光度(D)眼底检查透镜测量视盘尺寸时的观察者间一致性。
两名独立观察者使用结合了0.1毫米刻度标线的60 D透镜测量29只眼睛的视盘垂直直径和杯状凹陷垂直直径。计算垂直杯盘比。通过计算观察者间差异和加权kappa统计量评估观察者间一致性。
两名观察者对视盘直径测量(平均差异-0.04;范围-0.04, 0.2)和杯状凹陷直径测量(平均差异-0.03;范围0.3, 0.2)显示出良好的一致性。垂直杯盘比的一致性更高(平均kappa 0.96;95%置信区间0.90, 1.0)。杯盘比观察者间平均差异的95%置信区间为0.11,表明使用该技术评估杯盘比时变化>0.1在5%水平具有显著性。
使用该技术可实现视盘测量的高观察者间一致性。该方法有可能改善视盘的临床评估以及其他眼底结构临床测量的精度和准确性。