Higgins S L, Suh B D, Stein J B, Meyer D B, Jons J, Willis D
Regional Cardiac Arrhythmia Center, Scripps Memorial Hospital, La Jolla, CA 92037, USA.
Pacing Clin Electrophysiol. 1998 Jan;21(1 Pt 1):130-3. doi: 10.1111/j.1540-8159.1998.tb01072.x.
Twiddler's syndrome is a highly recognized yet rare complication of pacemaker and cardioverter defibrillator (i.c.d.) implantation. We present a case in which persistent generator rotation resulted in lead dislodgment and inappropriate shocks in an initial ICD and recurrent lead fracture in a second ICD system. This case is unusual in that even with extensive surgical precautions including use of a Dacron pouch, generator rotation could not be prevented. Submuscular implantation and use of a smaller generator may prevent Twiddler's syndrome.