Kammann J, Cosmar E, Walden K
St. Johannes Hospital, Dortmund, Germany.
J Cataract Refract Surg. 1998 Jan;24(1):98-106. doi: 10.1016/s0886-3350(98)80081-3.
To evaluate the clinical and visual results of the Medevec VS5 intraocular lens (IOL) with mini-loops, determine the incision size required for its implantation and its usable optical zone size, and compare the sizes with those of other foldable lenses.
Ophthalmology Department, St. Johannes Hospitals, Dortmund, Germany.
The VS5 lens was folded with a forceps and implanted into an intact bag in 27 patients; incision size was measured. Postoperatively, centration was recorded after dilation and the lens periphery observed to ascertain whether the loops were fixed in the capsular bag. Postoperative A-scans were done in 22 eyes, and the preoperative and postoperative vitreous cavity lengths were compared. The size of the optical zone was compared with that of other foldable lenses.
Forceps implantation of the VS5 lens required a 2.7 to 3.0 mm incision; all lenses were fixated in the bag. A-scans in 22 eyes showed that the vitreous cavity was shortened by an average of 0.51 mm. Uncorrected visual acuity was 20/40 or better in 96% of eyes. One week postoperatively, the IOLs remained centered and there was no flare or cells. The usable optical zone was larger than that of the staked long-loop lenses.
The Medevec VS5 fixated in the bag, consistently locating in the posterior bag space and stabilizing the vitreous. It had a full 5.0 mm optical zone and could be implanted through an incision smaller that that required by other foldable IOLs.
评估带有微型袢的Medevec VS5人工晶状体(IOL)的临床和视觉效果,确定其植入所需的切口大小和可用光学区大小,并将这些大小与其他可折叠晶状体的相应大小进行比较。
德国多特蒙德圣约翰医院眼科。
用镊子将VS5晶状体折叠后植入27例患者的完整囊袋中;测量切口大小。术后,散瞳后记录晶状体的中心定位情况,并观察晶状体周边以确定袢是否固定在囊袋中。对22只眼进行术后A超检查,比较术前和术后玻璃体腔长度。将光学区大小与其他可折叠晶状体的光学区大小进行比较。
用镊子植入VS5晶状体需要2.7至3.0毫米的切口;所有晶状体均固定在囊袋中。22只眼的A超检查显示玻璃体腔平均缩短了0.51毫米。96%的患眼未矫正视力达到20/40或更好。术后一周,人工晶状体保持居中,无闪光或细胞。可用光学区大于堆叠式长袢晶状体的光学区。
Medevec VS5固定在囊袋中,始终位于后囊袋空间并稳定玻璃体。它有完整的5.0毫米光学区,可通过比其他可折叠人工晶状体所需切口更小的切口植入。