Evaluation of a diode detector array for use as a linear accelerator QC device.

作者信息

Watts R J

机构信息

Live Oak Regional Cancer Center, San Antonio, Texas 78233, USA.

出版信息

Med Phys. 1998 Feb;25(2):247-50. doi: 10.1118/1.598188.

Abstract

A series of tests were conducted to evaluate the accuracy of a diode detector array in measuring flatness and symmetry of x-ray and electron energies, and the ability to determine change in beam energy. The results show the diode detector array accurately measures flatness and symmetry and can accurately measure or detect change in beam energy. The evaluation also demonstrates the utility of using this model detector system in a linear accelerator QC program.