新型超声胎儿心率监测仪准确性的评估

Evaluation of the accuracy of a new ultrasonic fetal heart rate monitor.

作者信息

Lauersen N H, Hochberg H M, George M E

出版信息

Am J Obstet Gynecol. 1976 Aug 15;125(8):1125-35. doi: 10.1016/0002-9378(76)90819-x.

Abstract

An indirect Doppler fetal monitoring system has been developed and validated by computer comparison of simultaneous fetal heart rate (FHR) with Doppler and scalp ECG of high-risk patients during labor. The difference in measurement of FHR averaged 0.3 b.p.m., and 93 per cent of the Doppler FHR measures were within 10 b.p.m. of the ECG FHR. If interinstrument difference is discounted, 96 per cent were within 10 b.p.m. All types of decelerations and variability were well approximated. Doppler FHR from the instrument described may be relied upon as valid clinical information and may be obtained from over 90 per cent of labor patients with 93 per cent accuracy.

摘要

一种间接多普勒胎儿监测系统已被开发出来,并通过计算机对高危患者分娩期间同时记录的多普勒胎儿心率(FHR)与头皮心电图进行比较来验证。FHR测量值的平均差异为0.3次/分钟,93%的多普勒FHR测量值与心电图FHR相差在10次/分钟以内。如果不考虑仪器间差异,96%的测量值相差在10次/分钟以内。所有类型的减速和变异性都能得到很好的近似。所描述仪器测得的多普勒FHR可作为有效的临床信息,且能从90%以上的分娩患者中获得,准确率达93%。