"Runaway" in a modern generation pacemaker.

作者信息

Odabashian H C, Brown D F

出版信息

Pacing Clin Electrophysiol. 1979 Mar;2(2):152-5. doi: 10.1111/j.1540-8159.1979.tb05194.x.

Abstract

Life-threatening episodes of ventricular tachycardia were documented in a patient with an implanted Omni-Stanicor pacemaker due to episodic pacemaker runaway. The unit generated impulses at rates varying from 73 to 500 per minute. Ventricular capture at high rates was variable, but frequently resulted in heart rates in excess of 200 per minute. These episodes were associated with loss of consciousness secondary to inadequate cardiac output. Although the pacemaker could be inactivated by reprogramming its output to 2 mA, and suppressed by chest wall stimulation with an external pulse generator, no spontaneous rhythm resulted during these maneuvers. Urgent removal of the battery at the bedside and immediate resumption of pacing with an external generator was required to resolve the problem. Although runaway is infrequent with current generation pacemakers, it should be considered in addition to the more common causes of pacemaker malfunction in the patient with appropriate symptoms.

摘要

相似文献

1
"Runaway" in a modern generation pacemaker.
Pacing Clin Electrophysiol. 1979 Mar;2(2):152-5. doi: 10.1111/j.1540-8159.1979.tb05194.x.
2
Runaway pacemaker: a still existing complication and therapeutic guidelines.
Clin Cardiol. 1989 Jul;12(7):412-4. doi: 10.1002/clc.4960120713.
3
Runaway pacemaker inhibited by external overdrive stimulation: a case report.
Pacing Clin Electrophysiol. 1984 May;7(3 Pt 1):440-2. doi: 10.1111/j.1540-8159.1984.tb04929.x.
4
Pacemaker runaway following intermittent output failure.
Pacing Clin Electrophysiol. 1983 May;6(3 Pt 1):645-7. doi: 10.1111/j.1540-8159.1983.tb05307.x.
5
Runaway pacemaker caused by a stuck accelerometer.
Pacing Clin Electrophysiol. 1999 Mar;22(3):538-9. doi: 10.1111/j.1540-8159.1999.tb00487.x.
6
[The case of the "runaway pacemaker"].
Kardiol Pol. 1991;35(8):103-5.
7
Runaway pacemaker.失控起搏器
BMJ Case Rep. 2019 Mar 31;12(3):e225411. doi: 10.1136/bcr-2018-225411.
9
Runaway pacemaker in seven patients: a persisting problem.
J Thorac Cardiovasc Surg. 1975 Mar;69(3):365-8.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验