Suppr超能文献

Runaway pacemaker: a still existing complication and therapeutic guidelines.

作者信息

Mickley H, Andersen C, Nielsen L H

机构信息

Department of Cardiology B, University Hospital, Odense, Denmark.

出版信息

Clin Cardiol. 1989 Jul;12(7):412-4. doi: 10.1002/clc.4960120713.

Abstract

Runaway pacemaker is a rare, but still existing potential lethal complication in permanent pacemakers. Within 4 1/2 years, we saw two cases of runaway pacemaker in patients with multiprogrammable, VVI pacemakers (Siemens-Elema, Model 668). In both cases a pacemaker-induced ventricular tachycardia (rate 240-260 beats/min) was documented. One patient died. Runaway pacemakers must be exchanged as soon as possible. Until this can be accomplished, different emergency maneuvers should be tried. As documented in the cases presented, placing a magnet over the pacemaker may result in a lower, more physiological pacing rate. Reprogramming the pulse generator to a lower output or the use of external chest wall overdrive stimulation may also be successful, but these procedures require the presence of an adequate escape rhythm. If this is not the case or the former maneuvers have failed, an external pacemaker may be connected to the permanent pacing lead. Thereafter, the lead can be safely cut. As an alternative, a temporary transvenous pacing lead may be established prior to disconnecting the permanent pacing lead.