Yokoyama M, Wada J, Hirosawa K
Pacing Clin Electrophysiol. 1979 Mar;2(2):225-9. doi: 10.1111/j.1540-8159.1979.tb05205.x.
This report presents the electrocardiograms of a patient with a normally functioning lithium QRS-inhibited pulse generator that exhibited partial recycling with inappropriately short escape intervals. The occurrence of partial recycling was time-dependent rather than voltage-dependent, and was observed only when QRS complexes fell within the relative refractory period of the pulse generator. This resulted in a complex arrhythmia simulating pulse generator malfunction. This observation re-emphasizes the need for the physician to appreciate the technical characteristics of pulse generators to avoid the unnecessary replacement of normally functioning units.