Falkoff M, Ong L S, Heinle R A, Barold S S
Pacing Clin Electrophysiol. 1978 Apr;1(2):250-3. doi: 10.1111/j.1540-8159.1978.tb03469.x.
Two patients with Omni-Stanicor pulse generators presented an apparent sensing problem characterized by intermittent reversion to fixed-rate pacing only during atrial fibrillation with a very rapid ventricular rate. Every fixed-rate cycle contained two unsensed beats. The first unsensed beat fell in the noise sampling period (the last 1/6 of the pacemaker refractory period) and, therefore, disabled the demand function of the pulse generator for a single timing cycle. The presence of two consecutively unsensed beats within one timing cycle (automatic or escape interval) during tachycardia suggests normal function of the noise sampling period of this particular pulse generator, rather than a true sensing problem. The diagnosis becomes evident if the sensing problem disappears when abbreviation of the refractory period occurs by reprogramming the pulse generator at a higher rate.
两名使用Omni-Stanicor脉冲发生器的患者出现了明显的感知问题,其特征为仅在房颤且心室率极快时间歇性恢复为固定频率起搏。每个固定频率周期包含两个未感知到的搏动。第一个未感知到的搏动落在噪声采样期(起搏器不应期的最后1/6)内,因此在单个计时周期内禁用了脉冲发生器的按需功能。心动过速期间在一个计时周期(自动或逸搏间期)内出现两个连续未感知到的搏动,提示该特定脉冲发生器的噪声采样期功能正常,而非真正的感知问题。如果通过将脉冲发生器编程为更高频率来缩短不应期时感知问题消失,则诊断变得明显。