Kind G M, Buntic R F, Buncke G M, Cooper T M, Siko P P, Buncke H J
Microsurgical Replantation Transplantation Department of Daives Medical Center, San Francisco, Calif 94114, USA.
Plast Reconstr Surg. 1998 Apr;101(5):1268-73; discussion 1274-5.
One hundred forty-seven flaps in 135 consecutive patients undergoing microvascular transplantation were monitored using a miniature Doppler ultrasonic probe. Using a modification of a technique described previously by Swartz, the probes were secured to the outflow vein of the flap with Vicryl mesh. Twenty instances of thrombosis or spasm were detected in 16 patients, and all flaps were salvaged (100 percent). There were four false positive and no false negative results. This probe allows for safe, continuous monitoring of flap blood flow, which permits the rapid detection and hence rapid treatment of postoperative complications. Our experience suggests that a significant improvement in the salvage rate of microvascular transplants may be attainable with the use of this device.
在135例连续接受微血管移植的患者中,使用微型多普勒超声探头对147个皮瓣进行了监测。采用对Swartz先前描述技术的改良方法,用薇乔网将探头固定在皮瓣的流出静脉上。在16例患者中检测到20例血栓形成或痉挛,所有皮瓣均得以挽救(100%)。有4例假阳性结果,无假阴性结果。该探头可对皮瓣血流进行安全、连续的监测,从而能够快速检测并因此快速治疗术后并发症。我们的经验表明,使用该设备可能会显著提高微血管移植的挽救率。