Cantor A, Goldfarb B, Aszodi A, Battler A
Cardiology Division, Soroka Medical Center, Faculty of Health Sciences, Ben-Gurion University of the Negev, Beer-Sheva, Israel.
J Electrocardiol. 1998 Jan;31(1):9-15. doi: 10.1016/s0022-0736(98)90002-4.
A new computerized optical scanner was used to measure QRS complex duration during exercise stress testing, both pre- and postdischarge, as a means of ischemia detection after acute myocardial infarction. Thallium stress testing was used as a standard of comparison. Each patient underwent predischarge exercise testing (while receiving anti-ischemic drug therapy) and a postdischarge test 1 month later (without anti-ischemic drug therapy), as well as thallium stress testing within 4 months of infarction. In the population of 68 patients, 42 of the predischarge tests and 43 of the postdischarge tests showed an ischemic response of QRS prolongation. When compared with thallium testing for QRS prolongation criteria, the sensitivity was 95% with a specificity of 77% predischarge and 89% with a specificity of 65% postdischarge. According to ST-T criteria, only 12 of 68 patients were positive for ischemia predischarge; this number increased to 29 postdischarge (predischarge sensitivity 24% and specificity 90%, with postdischarge sensitivity 68% and specificity 87%), when compared with thallium testing. Measuring QRS duration during exercise increased the sensitivity of detection of ischemic patients over that of ST-T criteria by 71% predischarge and 21% postdischarge, with a 22-23% loss of specificity, and was apparently not influenced by anti-ischemic drug therapy.
一种新型计算机光学扫描仪用于在运动负荷试验期间测量急性心肌梗死后出院前和出院后的QRS波群时限,作为缺血检测的一种手段。铊负荷试验用作对照标准。每位患者均接受出院前运动试验(接受抗缺血药物治疗时)、1个月后的出院后试验(不接受抗缺血药物治疗)以及梗死后4个月内的铊负荷试验。在68例患者群体中,42例出院前试验和43例出院后试验显示QRS波群延长的缺血反应。与铊试验的QRS波群延长标准相比,出院前敏感性为95%,特异性为77%;出院后敏感性为89%,特异性为65%。根据ST-T标准,68例患者中仅12例出院前缺血呈阳性;与铊试验相比,出院后这一数字增至29例(出院前敏感性24%,特异性90%;出院后敏感性68%,特异性87%)。运动期间测量QRS波群时限使缺血患者检测的敏感性在出院前比ST-T标准提高71%,出院后提高21%,特异性损失22%-23%,且显然不受抗缺血药物治疗的影响。