Wang L H, Wang J K, Mullins C E
Department of Pediatrics, National Taiwan University Hospital, Taipei.
Cathet Cardiovasc Diagn. 1998 Mar;43(3):295-7. doi: 10.1002/(sici)1097-0304(199803)43:3<295::aid-ccd11>3.0.co;2-l.
Two patients who underwent transcatheter closure of patent ductus arteriosus, one with a Rashkind umbrella device and the other with a coil, suffered from acute hemolysis following the procedure. Hemolysis ceased after deployment of second device(s) within 48 hr without needing to retrieve the first devices in either patient. We conclude that immediate deployment of a second device(s) is an alternative to surgery when acute hemolysis occurs following transcatheter closure of ductus.
两名接受动脉导管未闭经导管封堵术的患者,其中一名使用Rashkind伞装置,另一名使用弹簧圈,术后均出现急性溶血。两名患者在48小时内植入第二个装置后溶血停止,无需取出第一个装置。我们得出结论,当动脉导管经导管封堵术后发生急性溶血时,立即植入第二个装置是手术的一种替代方法。