We report on two latex-allergic subjects who developed local hypersensitivity reactions after their lung compliance was measured by an oesophagus balloon catheter. Protein and allergen analyses of this catheter showed values of 85 and 9.8 micrograms per gram rubber. It is concluded that latex-allergic subjects should not undergo examinations with latex allergen-containing catheters. To our knowledge, this is the first report on allergic reactions caused by latex balloons of an oesophagus catheter.
我们报告了两名对乳胶过敏的受试者,他们在用食管气囊导管测量肺顺应性后出现了局部过敏反应。对该导管进行的蛋白质和过敏原分析显示,每克橡胶中的含量分别为85微克和9.8微克。得出的结论是,对乳胶过敏的受试者不应使用含有乳胶过敏原的导管进行检查。据我们所知,这是关于食管导管乳胶气囊引起过敏反应的首例报告。