[Evaluation of patient exposure to radiation during x-ray diagnostic examinations (in selected health service centers in the city of Wrocław and in the province of Wrocław)].

作者信息

Demczuk P, Jankowski J

机构信息

Oddziału Ochrony Radiologicznej, Wojewódzkiej Stacji Sanitarno-Epidemiologicznej we Wrocławiu.

出版信息

Med Pr. 1997;48(6):639-50.

PMID:9558632
Abstract

In 35 x-ray laboratories, randomly selected, in the city of Wrocław an in the voivodship, the following measurements were performed: -exposure doses (mGy) for different kinds of x-ray examinations, -exposure doses (mGy) for examinations employing conventional intensifying cassettes and cassettes equipped with intensifying screens of rare earths, -basic exposure parameters, ie high voltage (kV), anodic current intensity (mA), and exposure duration. An average number of exposure per one examination was also defined. The values of standard deviation (SD) for doses measured and for exposure parameters were identified. The doses were compared with the values recommended in "Basic Safety Standards for Protection Against Ionizing Radiation" (IAEA, Safety Series No 115). Two differences were discussed between the values measured and those recommended, particularly in exposures when conventional cassettes were used. The assessment of exposure parameters for the same x-ray examinations performed in different laboratories pointed to a lack of any analogy between the results. It bears witness to the fact that x-ray equipment is very much diversified, and that subjective factors play a significant role.