Haeckel R
J Clin Chem Clin Biochem. 1976 May;14(5):227-37. doi: 10.1515/cclm.1976.14.1-12.227.
This report summarizes a one year evaluation of Abbott's ABA 100, with respect to mechanical parts (syringe plates, precision and linearity of photometry, band width of several filters, multicuvet precision, temperature control) and the reliability of several methods (endpoint procedures: determination of the glucose concentration with hexokinase- and the glucose dehydrogenase method, and of the protein concentration; enzyme activities: alanine and aspartate aminotransferase, creatine kinase, alkaline phosphatase). The critical batch size was estimated as an indicator of economy (about 40 samples per day for the glucose concentration). Various aspects of the instrument are discussed with respect to its use in clinical chemistry.
本报告总结了对雅培ABA 100进行的为期一年的评估,评估内容涉及机械部件(注射器板、光度测定的精度和线性、多个滤光片的带宽、多比色皿精度、温度控制)以及几种方法的可靠性(终点法:用己糖激酶法和葡萄糖脱氢酶法测定葡萄糖浓度以及蛋白质浓度;酶活性:丙氨酸氨基转移酶、天冬氨酸氨基转移酶、肌酸激酶、碱性磷酸酶)。估算了关键批量大小作为经济性指标(葡萄糖浓度测定每天约40个样本)。针对该仪器在临床化学中的应用,讨论了其各个方面。