使用近红外和红光容积脉搏波描记法测量手指小动脉容积脉搏在不同倾斜程度下的变化。

Changes in the arteriolar volume pulse of the finger during various degrees of tilt using near infra-red and red photoplethysmography.

作者信息

Christ F, Nehring I, Abicht J, Baranov V, Kotov A, Gartside I, Gamble J, Messmer K

机构信息

Institut für Anaesthesiologie, Ludwig Maximillians University Munich, Klinikum Grosshadern, Marchioninistr. 15, D-82362 Munich, Germany.

出版信息

Eur J Med Res. 1998 May 12;3(5):249-55.

PMID:9580571
Abstract
  1. Photoplethysmography is a widely used non invasive technique for the measurement of peripheral oxygen saturation. A more detailed analysis of the volume pulse (VP) can indicate alterations in peripheral vascular tone, due to sympathetic stimulation, stress, pain and temperature. - 2. In six healthy male volunteers we investigated changes in the VP resulting from vasoconstriction and vasodilatation induced by varies degrees of tilt. Subjects were subjected 0 degrees tilt followed by head down -8 degrees -15 degrees, -30 degrees -15 degrees, -8 degrees, 0 degrees, 15 degrees, 30 degrees , 70 degrees, 30 degrees, 15 degrees and 0 degrees. Each tilt stage was sustained for 15 minutes. Both VP - and haemodynamic changes were continuously recorded 30s before and then for 210 s after the imposition of each tilt step. We used a new computer driven soft and hardware for the analysis of the VP. 3. The VP signal was obtained with a sensor emitting 840 nm (NIR) and 640 nm (RED) light into finger tip with a sample rate of 128 Hz. All data was normalised to the initial mean value obtained at 0 degrees tilt. The signal strength parameters amplitude, and area under the curve and the first derivative of the amplitude (flux) as well as time discrete parameters, time of first maximum (Tmax), dicrote wave (Td), volume pulse decrease (Tdec) and fundamental arterial oscillation Tag = Td - Tmax were measured. 4. HR increased significantly during 30 degrees and 70 degrees tilt, but no change in the other hemodynamic parameters was observed. Amplitude, area under the curve and flux of both the Red and the NIR signal increased following head down tilt. A significant decrease of those parameters was found during foot down tilt. 5. No significant changes were found in the time discrete values, neither within each tilt step nor when compared to the initial mean value at 0 degrees tilt. 6. This study reveals that signal strength related parameters such as area under the curve, amplitude and flux reflect changes in vascular tone. Time discrete parameters however did not depict these changes and appear unsuitable for data analysis when using this specific hardware applied in the current study.
摘要
  1. 光电容积脉搏波描记法是一种广泛应用于测量外周血氧饱和度的非侵入性技术。对容积脉搏(VP)进行更详细的分析可以显示由于交感神经刺激、压力、疼痛和温度引起的外周血管张力变化。

  2. 在六名健康男性志愿者中,我们研究了不同程度倾斜诱导的血管收缩和舒张所导致的VP变化。受试者先处于0度倾斜,然后依次为头向下-8度、-15度、-30度、-15度、-8度、0度、15度、30度、70度、30度、15度和0度。每个倾斜阶段持续15分钟。在每次施加倾斜步骤前30秒及之后210秒持续记录VP和血流动力学变化。我们使用了一种新的计算机驱动的软硬件来分析VP。

  3. 通过一个向指尖发射840纳米(近红外)和640纳米(红光)光的传感器以128赫兹的采样率获取VP信号。所有数据均根据在0度倾斜时获得的初始平均值进行归一化。测量信号强度参数振幅、曲线下面积以及振幅的一阶导数(通量),以及时间离散参数,首次最大值时间(Tmax)、重搏波(Td)、容积脉搏下降(Tdec)和基本动脉振荡Tag = Td - Tmax。

  4. 在30度和70度倾斜期间心率显著增加,但未观察到其他血流动力学参数的变化。头向下倾斜后,红光和近红外信号的振幅、曲线下面积和通量均增加。在脚向下倾斜期间发现这些参数显著下降。

  5. 在时间离散值方面,无论是在每个倾斜步骤内还是与0度倾斜时的初始平均值相比,均未发现显著变化。

  6. 本研究表明,诸如曲线下面积、振幅和通量等与信号强度相关的参数反映了血管张力的变化。然而,时间离散参数并未描述这些变化,并且在使用本研究中应用的这种特定硬件时似乎不适用于数据分析。