van der Laak J A, Westphal J R, Schalkwijk L J, Pahlplatz M M, Ruiter D J, de Waal R M, de Wilde P C
Institute of Pathology, University Hospital Nijmegen, The Netherlands.
J Pathol. 1998 Feb;184(2):136-43. doi: 10.1002/(SICI)1096-9896(199802)184:2<136::AID-PATH970>3.0.CO;2-9.
In a number of recent papers, the degree of tumour vascularization has been described as a promising new prognostic factor. Methods for the assessment of vascular density involve immunohistochemical staining of the vasculature, followed by counting the number of vessel profiles in the angiogenic hot spot. One of the problems of this procedure is the selection of the angiogenic hot spot, which has been described as being subject to inter-observer variation. In this study, the value of true colour image analysis in reducing inter-observer variation has been assessed. Highly (MV3) and poorly (M14) vascularized human melanoma xenografts were used to evaluate the image analysis procedure, and the image analysis results were compared with results from the conventional manual hot-spot procedure. Assessment by image analysis was performed on measurement fields covering the entire tumour tissue specimens rather than on a single hot-spot field. Also, by selecting the most densely vascularized area from all fields assessed by the semi-automatic procedure, it was possible to objectify the hot spot selection (automated hot-spot procedure). Manual assessment showed a good correlation between two independent observers for MV3 xenografts (r = 0.74, P = 0.014), but a poor correlation for M14 xenographs (r = 0.4, P > 0.05). Automated assessment by different operators showed good correlations for both MV3 xenografts (r = 0.99, P < 0.001) and M14 xenografts (r = 0.80, P = 0.006). It is concluded that although both manual vessel counting and semi-automated image analysis can differentiate between the level of vascularization in the two types of xenograft (P < 0.001 for both methods), the automated method is favourable in that it showed no significant inter-observer effects. In M14 xenografts, the manual hot-spot vessel densities did not correlate well with the automated hot-spot densities (r = 0.27, P > 0.05), indicating that selection of angiogenic hot spots in this tumour type is indeed subject to observer bias. The automated hot-spot vessel densities were a reliable indicator of overall tumour vessel density in both tumour types. Image analysis allows analysis of vessel subclasses based on morphological criteria such as vessel profile area or diameter. In the model system used, the discrimination between MV3 and M14 xenografts was further enhanced by selectively examining vessels with diameters between 6 and 9 microns (P < 0.0005). In conclusion, image analysis appears to offer an objective and more reproducible method to quantify tumour vascularity than manual counting of vessel profiles in the hot spot. Analysis of subclasses of vessels may further enhance the value of vessel density measurements in discriminating between tumour types differing in biological behaviour.
在最近的一些论文中,肿瘤血管化程度已被描述为一种很有前景的新预后因素。评估血管密度的方法包括对脉管系统进行免疫组织化学染色,然后在血管生成热点区域计数血管轮廓的数量。该程序的问题之一是血管生成热点的选择,这已被描述为存在观察者间差异。在本研究中,评估了真彩色图像分析在减少观察者间差异方面的价值。使用高度血管化(MV3)和低血管化(M14)的人黑色素瘤异种移植模型来评估图像分析程序,并将图像分析结果与传统的手动热点程序结果进行比较。通过图像分析进行的评估是在覆盖整个肿瘤组织标本的测量区域上进行的,而不是在单个热点区域上。此外,通过从半自动程序评估的所有区域中选择血管最密集的区域,可以使热点选择客观化(自动热点程序)。手动评估显示,对于MV3异种移植模型,两名独立观察者之间具有良好的相关性(r = 0.74,P = 0.014),但对于M14异种移植模型,相关性较差(r = 0.4,P>0.05)。不同操作人员的自动评估显示,对于MV3异种移植模型(r = 0.99,P <0.001)和M14异种移植模型(r = 0.80,P = 0.006)均具有良好的相关性。得出的结论是,虽然手动血管计数和半自动图像分析都可以区分两种异种移植模型中的血管化水平(两种方法的P均<0.001),但自动方法更具优势,因为它没有显示出明显的观察者间效应。在M14异种移植模型中,手动热点血管密度与自动热点密度之间的相关性不佳(r = 0.27,P>0.05),这表明在这种肿瘤类型中血管生成热点的选择确实存在观察者偏差。自动热点血管密度是两种肿瘤类型中总体肿瘤血管密度的可靠指标。图像分析允许根据血管轮廓面积或直径等形态学标准分析血管亚类。在所使用的模型系统中,通过选择性检查直径在6至9微米之间的血管,进一步增强了MV3和M14异种移植模型之间的区分度(P <0.0005)。总之,与在热点区域手动计数血管轮廓相比,图像分析似乎提供了一种更客观、更可重复的方法来量化肿瘤血管生成。分析血管亚类可能会进一步提高血管密度测量在区分生物学行为不同的肿瘤类型方面的价值。